Safe Diffusion Lobes for Diminished Gradient Vibration
Lists
Only you see this
Abstract
Motivation: Gradient coil vibrations are a source of loud acoustics, persistent field distortions, potential system damage, and in some cases signal dropout in diffusion MRI. Vibrations are particularly strong at mechanical resonant frequencies. Goal(s): To design “safe” time-optimal diffusion gradients which avoid mechanical resonant frequencies. Approach: A convex optimization problem is formed which seeks to maximize b-value for a given duration while constraining the discrete cosine transform of the diffusion lobe. A bisection search is applied to duration to find the shortest feasible solution with the target b-value. Results: Safe diffusion lobes greatly diminished coil vibration and minimized oscillations in the gradient fields. Impact: Safe diffusion lobes were shown to avoid specified gradient mechanical resonant frequencies reducing vibrations and persistent k-space oscillations. These waveforms have clinical and research potential to improve auditory comfort, prevent DWI signal dropout and image artifacts, and prevent system damage.
BibTeX
@inproceedings{mccready2024safe,
author = {Mark J. McCready and John M. Pauly and Adam B. Kerr},
title = {{Safe Diffusion Lobes for Diminished Gradient Vibration}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},
doi = {10.58530/2024/0912},
}Loading related content...