NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

Retrospective motion gating in cardiac MRI using a simultaneously acquired navigator

Byungjai Kim, Hyunseok Seo, Dongchan Kim, HyunWook ParkNMR 2017

Abstract

A simultaneous acquisition technique of image and navigator signals (simultaneously acquired navigator, SIMNAV) is proposed for cardiac magnetic resonance imaging (CMRI) in Cartesian coordinates. To simultaneously acquire both image and navigator signals, a conventional balanced steady-state free precession (bSSFP) pulse sequence is modified by adding a radiofrequency (RF) pulse, which excites a supplementary slice for the navigator signal. Alternating phases of the RF pulses make it easy to separate the simultaneously acquired magnetic resonance data into image and navigator signals. The navigator signals of the proposed SIMNAV were compared with those of current gating devices and self-gating techniques for seven healthy subjects. In vivo experiments demonstrated that SIMNAV could provide cardiac cine images with sufficient image quality, similar to those from electrocardiogram (ECG) gating with breath-hold. SIMNAV can be used to acquire a cardiac cine image without requiring an ECG device and breath-hold, whilst maintaining feasible imaging time efficiency.
BibTeX
@inproceedings{kim2017retrospective,
  author = {Byungjai Kim and Hyunseok Seo and Dongchan Kim and HyunWook Park},
  title = {{Retrospective motion gating in cardiac MRI using a simultaneously acquired navigator}},
  booktitle = {NMR},
  year = {2017},
  doi = {10.1002/nbm.3874},
}

Loading related content...

Figure