Research on measurement traceability of small force values based on the principle of electromagnetic force balance

Lifang Wang, Jing Yuan, Jianfeng Zhang, Ying Wu, Yiwen Tang2025

Abstract

The micro force calibration device achieves nanoscale displacement control through a spiral differential cylinder and piezoelectric ceramics. Based on the principle of electromagnetic force balance, it accurately measures the coil current to achieve precise measurement of small external forces. The device has a measurement uncertainty of <i>U<sub>rel</sub></i> = 0.2% ~ 0.6%(<i>k</i> = 2) within the range of 10 mN&lt;F&le;100 mN; The measurement uncertainty is <i>U<sub>rel</sub></i> = 0.2% ~ 0.6%(<i>k</i> = 2) within the range of 100 mN&lt;F&le;5N. This article conducted micro force measurement experiments under periodic environmental temperature loads and found that the micro force calibration device is affected by temperature to -0.06199 N/&deg;C. Under constant temperature laboratory conditions, the verification test of the micro force calibration device and the U1 dynamometer with a 0.02 level HBM was completed. The difference in the measurement results of the reference line slope K between the two was 0.000275 (mV/V)/N.
BibTeX
@misc{wang2025research,
  author = {Lifang Wang and Jing Yuan and Jianfeng Zhang and Ying Wu and Yiwen Tang},
  title = {{Research on measurement traceability of small force values based on the principle of electromagnetic force balance}},
  year = {2025},
  doi = {10.1117/12.3072968},
}

Loading related content...