Reduction of metallic artifacts from the novel leadless pacemaker using single-energy metal artifact reduction algorithm in coronary CT angiography

Yahui Hu, Ying Chen, Jianjun Huo, Linxin Zhang, Yan Li, Qian Liu, Yaohui Yu, Haicheng Qi, Zicheng Zhao, Jiaojiao Zhang, Min Xu, Yan XingBMC Medical Imaging 2025 · DOI

Abstract

Metal artifacts from implanted cardiac devices significantly hinder the diagnostic utility of Coronary CT Angiography (CCTA). This study investigates the performance of the Single-Energy Metal Artifact Reduction (SEMAR) algorithm in patients with novel leadless pacemakers (PMs) versus traditional transvenous PMs. Analyzing 60 clinical cases, the study finds that SEMAR substantially mitigates beam-hardening and photon-starvation effects. Leadless PMs, due to their compact design, were found to interfere less with critical coronary anatomy, and when combined with SEMAR, provided superior diagnostic confidence compared to transvenous alternatives.

Introduction

Pacemakers (PMs) are essential for managing cardiac rhythm disorders, yet their metallic composition poses a significant challenge for medical imaging. In Coronary CT Angiography (CCTA), high-density metal components cause beam-hardening and photon-starvation, resulting in dark shadows and bright streaks known as artifacts. These artifacts can obscure coronary arteries and the myocardium, potentially masking conditions like coronary artery disease or device-related complications.

While hardware solutions like Dual-Energy CT exist, they are not universally available. Software-based solutions, specifically the Single-Energy Metal Artifact Reduction (SEMAR) algorithm, offer a more accessible alternative for standard CT scanners. This study specifically evaluates SEMAR's effectiveness on the newer generation of leadless pacemakers—compact devices implanted directly into the right ventricle—compared to traditional transvenous models.

Methods

The researchers conducted a retrospective study involving 60 patients (30 with leadless PMs and 30 with transvenous PMs) who underwent CCTA. Raw CT data were reconstructed using two methods: standard Hybrid-Iterative Reconstruction (HIR) as a baseline, and HIR integrated with the SEMAR algorithm.

The evaluation was both quantitative and qualitative:

  1. Quantitative Analysis: An Artifact Index (AI) was calculated by measuring CT attenuation (Hounsfield Units, HU) within Regions of Interest (ROIs) placed in the most severe artifact zones compared to a reference background of clean tissue.
  2. Qualitative Analysis: Two senior radiologists independently scored the images on a 5-point scale for artifact severity and diagnostic confidence. They also assessed the visibility of specific anatomical structures, such as the aorta and coronary arteries.
Figure 5: Representative CT image showing the placement of Regions of Interest (ROIs) used to calculate the Artifact Index in the presence of pacemaker-induced metal artifacts.

Results

The application of SEMAR led to a dramatic improvement in image quality across both device types, with leadless PMs showing the most favorable outcomes.

Quantitative Improvements

For leadless PMs, the Artifact Index (AI) decreased from 243.82 HU in baseline reconstructions to 59.28 HU with SEMAR ($p < 0.001$). This significant reduction indicates a substantial clearing of the "noise" caused by the metal components.

Qualitative and Diagnostic Gains

Radiologists reported a marked increase in diagnostic confidence. For leadless PM cases, confidence scores rose from a restricted 2.20/5 to a high 4.62/5 after SEMAR processing.

Anatomical Sparing

A key finding was that leadless PMs naturally obscure less critical anatomy. Even without SEMAR, over 80% of Right Coronary Artery (RCA) evaluations were rated as "unimpaired" for leadless devices, compared to only 20% for transvenous devices. The compact form factor of leadless PMs prevents the widespread artifact distribution seen with the long wires of transvenous systems.

Discussion and Conclusion

The study demonstrates that the SEMAR algorithm is a highly effective, software-based tool for improving CCTA image quality in patients with pacemakers. The combination of leadless PM technology and SEMAR reconstruction offers a significant advantage for cardiac imaging, preserving the visibility of the RCA and other vital structures.

While SEMAR is effective, the authors noted it could occasionally introduce minor secondary artifacts in highly degraded areas. Future research should involve larger, multi-center cohorts and focus on whether these improvements in image quality translate directly into better clinical diagnostic accuracy for pacemaker-related complications.

Summary updated