Abstract

Magnetic resonance imaging (MRI) in the presence of metallic objects suffers from slice-selection errors and in-plane distortions. View angle tilting (VAT) corrects for in-plane distortions by adding a gradient on the slice-select axis during readout, but can suffer from image blurring. This work demonstrates that the major source of blurring is a slice profile modulation of the data, and proposes several solutions to prevent such blurring. Multiple high-bandwidth readouts are demonstrated to reduce the blurring while improving the signal-to-noise ratio (SNR) over a single high-bandwidth readout.