Abstract
Motivation: BLADE (PROPELLER) is robust to motion but more prone to aliasing-induced artifacts. Goal(s): To minimize aliasing-induced artifacts in BLADE(PROPELLER) sequence. Approach: Utilizing GRAPPA method to extend FOV for each blade in reconstruction. Pre-scan data is used to calibrate the GRAPPA kernels for each blade. Results: Proposed method can effectively eliminate aliasing-induced artifacts. Impact: The extended-FOV method can be integrated seamlessly into the existing scanning process without requiring additional scan time, to effectively eliminate aliasing-induced artifacts.
BibTeX
@inproceedings{zhou2025reducing,
author = {Kun Zhou and Nan Xiao and Yang Li},
title = {{Reducing aliasing-induced artifacts in BLADE sequence using extended-FOV reconstruction}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2025},
doi = {10.58530/2025/5134},