Rapid Separation and Determination of Amino Acids by High-performance Liquid Chromatography on a Xterra RP18 Column
Abstract
A method for rapid separation and determination of amino acids by high performance liquid chromatography on a Xterra RP18 column was established. The chromatogram conditions were: the column temperature of 56℃, the flow rate of 1.8 ml/min and the detecting wavelength of 248nm; gradient separation on the Waters Alliance TM system, the run time of 25 minutes and the counterpressure of chromatogram column under 2000 psi. 21 amide was separated within 17.5 minutes, and AMQ, NH 3 and taurine were included. The method could be used for the determination of amino acids in compound amino acid injection liquid, oral amino acid liquid including taurine and compound amino acid hydrolyzed sample.
BibTeX
@inproceedings{chen2001rapid,
author = {Yongbo Chen and Tang Deng-mei},
title = {{Rapid Separation and Determination of Amino Acids by High-performance Liquid Chromatography on a Xterra RP18 Column}},
booktitle = {Amino},
year = {2001},