Abstract
Routine tongue movement therapy with nerve tip stimulation serves as a critical therapeutic aid for patients suffering from stroke, dysphagia, and sleep apnea. This research details the design and construction of a Tongue Pressure Training (LTL) system. The system utilizes Force Sensing Resistor (FSR) sensors to facilitate and monitor three specific tongue exercises: Tongue Force, Tongue Slide, and Tongue Press. Data is processed via an Arduino microcontroller and transmitted via Bluetooth to a dedicated smartphone application for user feedback. Experimental results demonstrate that the LTL system successfully measures tongue pressure in the range of 0–730 Gram-force. While a 1-second average latency exists between sensor reading and display, the system effectively differentiates exercise types through spatial sensor placement.
Introduction
Conditions such as stroke and dysphagia (difficulty swallowing) often stem from impaired motor nerves in the mouth and throat. Similarly, sleep apnea is frequently associated with weakened tongue muscles that can obstruct airways during sleep. Tongue exercises are a proven intervention to stimulate neural pathways and strengthen these muscles. However, clinical-grade monitoring equipment is often prohibitively expensive or restricted to hospital environments. This study addresses the need for an affordable, portable, and user-friendly device that allows patients to perform guided exercises and track their progress quantitatively at home.
System Architecture and Methodology
The LTL system is built upon a modular hardware architecture designed for portability and ease of use. The core components include:
- Sensors: Two Interlink FSR-402 sensors are employed to detect pressure. These sensors vary their resistance based on the force applied, which is then converted into physical units (Gram-force).
- Processing and Communication: An Arduino Uno serves as the central processing unit, reading analog signals from the FSRs. A Bluetooth HC-06 module facilitates wireless data transmission to a mobile device.
- Software Interface: A smartphone application, developed using MIT App Inventor, provides a user interface for identity input, exercise selection, and real-time result visualization.
Sensor Placement and Exercise Mechanics
A key innovation of the LTL system is the strategic placement of sensors on a custom mouthpiece to distinguish between different motor patterns.
As shown in the mouthpiece design, Sensor 1 is positioned at the front to capture tip pressure, while Sensor 2 is placed further back. This configuration allows the system to identify specific "exercise signatures":
- Tongue Press and Tongue Force: These movements primarily activate Sensor 1, with pressure readings reaching up to approximately 640 gf.
- Tongue Slide: This movement involves the tongue moving backward across the mouthpiece, creating a sequential pressure profile across both Sensor 1 and Sensor 2. This exercise recorded the highest pressures, up to 730 gf on the primary sensor.
Results and Discussion
The prototype was evaluated through a series of calibration and latency trials. The system successfully mapped non-linear sensor resistance to a linear-like scale of 0–730 Gram-force, providing a concrete metric for rehabilitation progress.
One technical challenge identified during testing was system latency. There is an average delay of 1 second between the physical tongue movement and the update on the smartphone interface. While this does not prevent the recording of results, it is a bottleneck for real-time interactive feedback. Despite this, the system proved highly effective at distinguishing the three target movements and providing a reliable history of exercise performance.
Conclusion and Future Work
The LTL system successfully demonstrates the feasibility of a low-cost, smartphone-integrated solution for tongue rehabilitation. By shifting the locus of care from the clinic to the home, such devices can significantly improve patient compliance and recovery outcomes.
Future development will focus on reducing system latency to provide more instantaneous feedback and enhancing the smartphone application's UI to be more engaging. Most importantly, the next phase of research will involve clinical validation with stroke and dysphagia patients to refine mouthpiece ergonomics and therapeutic efficacy.
Summary updated