Radiosynthesis automation, non-human primate biodistribution and dosimetry of K + channel tracer [ 11 C]3MeO4AP

Yu‐Peng Zhou, Moses Q. Wilks, Maëva Dhaynaut, Nicolas J. Guehl, Sung‐Hyun Moon, Georges El Fakhri, Marc D. Normandin, Pedro BrugarolasbioRxiv 2023

TL;DR

This paper presents a fully automated, cGMP-compliant radiosynthesis of a novel PET tracer, [¹¹C]3MeO4AP, designed to image potassium channels. Biodistribution and dosimetry studies in non-human primates show promising results, indicating efficient brain uptake and a favorable radiation dose profile compared to similar tracers, paving the way for clinical applications in neurological disorders like multiple sclerosis.

What Problem Does This Paper Address?

Multiple sclerosis (MS) and other demyelinating diseases involve the exposure and increased expression of axonal potassium channels (Kv1.1 and Kv1.2) as the myelin sheath degrades. 4-Aminopyridine (4AP) is a medication that blocks these channels, restoring axonal conduction. To better understand and diagnose these conditions, researchers are developing PET tracers based on 4AP that can visualize these potassium channels in vivo. While a fluorine-18 labeled tracer ([¹⁸F]3F4AP) has shown promise, carbon-11 (¹¹C) tracers offer advantages like shorter half-lives, leading to lower radiation doses and the possibility of same-day repeat scans. However, developing ¹¹C tracers requires efficient, automated radiosynthesis under Good Manufacturing Practices (cGMP) for clinical translation, along with thorough biodistribution and dosimetry studies to ensure safety. This paper addresses the need for a cGMP-compliant automated synthesis of a specific ¹¹C-labeled 4AP analog, [¹¹C]3MeO4AP, and evaluates its suitability for human use through preclinical studies in non-human primates.

Methods

The core of this work involved developing and validating a fully automated radiosynthesis of [¹¹C]3MeO4AP using a GE TRACERlab FX C Pro synthesizer. This automated process was designed to meet cGMP standards, ensuring reproducibility and minimizing operator radiation exposure. The synthesis involved the radiomethylation of 3-hydroxyl-4-aminopyridine with [¹¹C]CH₃I, which was generated from [¹¹C]CO₂.

Following radiosynthesis, the tracer's quality was assessed using a battery of tests, including visual inspection, radio-HPLC for radiochemical identity and purity, dose calibrator and HPLC for molar activity, and half-life measurement for radionuclidic identity.

For biodistribution and dosimetry, two non-human primates (one male, one female rhesus macaque) were injected with the tracer. PET/CT imaging was performed dynamically for approximately 150 minutes. Volumes of interest (VOIs) were manually drawn on organs from the CT and PET images. Time-activity curves (TACs) were extracted for each VOI. Blood samples were also collected serially to assess blood clearance. Radiation dosimetry and effective dose were calculated using the OLINDA software, integrating the TACs and extrapolating to ten half-lives.

Datasets

The study utilized data from two non-human primates (one male, one female rhesus macaque) for biodistribution and dosimetry evaluation. No new benchmark datasets were introduced.

Key Insights and Ideas

  • cGMP Automation is Key for Clinical Translation: The development of a fully automated radiosynthesis process that adheres to cGMP standards is a critical step for moving novel PET tracers from preclinical research to human clinical trials. This not only ensures product quality and safety but also minimizes radiation exposure for the radiochemist.
  • [¹¹C]3MeO4AP Shows Promising Brain Uptake and Renal Clearance: The biodistribution data revealed that [¹¹C]3MeO4AP distributes widely throughout the body, with significant uptake in the brain. The rapid clearance from most organs, particularly the kidneys and urinary bladder, suggests efficient renal excretion, which is generally favorable for reducing radiation burden to other tissues.
  • Favorable Radiation Dosimetry: The calculated effective dose for [¹¹C]3MeO4AP was found to be significantly lower than its ¹⁸F analog ([¹⁸F]3F4AP) and comparable to other widely used ¹¹C tracers like [¹¹C]choline and [¹¹C]glucose. This low effective dose is a major advantage for repeated imaging or use in patient populations.
  • No Significant Sex-Based Differences: The study found no significant differences in biodistribution or dosimetry between the male and female non-human primates, simplifying potential future human studies.

Key Results

  • Automated Radiosynthesis: A fully automated radiosynthesis of [¹¹C]3MeO4AP was successfully developed, achieving a radiochemical yield of 7.3 ± 1.2% and >99% radiochemical purity within 38 minutes. This is a 3% lower yield than the semi-automated method but sufficient for clinical doses and achieved in a shorter time.
  • Biodistribution: [¹¹C]3MeO4AP showed widespread distribution, with accumulation in the urinary bladder, kidneys, liver, thyroid, brain, salivary glands, and spinal vertebrae. Kidneys and urinary bladder showed the highest tracer accumulation, indicating primary renal clearance.
  • Blood Clearance: Blood clearance was rapid, with a two-phase decay model indicating fast washout (t½ fast ≈ 0.5-0.7 min) and slower washout (t½ slow ≈ 9-19 min) depending on sex.
  • Brain Kinetics: Brain uptake was slower than in other organs, reaching a maximum SUV of ~2.4 at 16-29 minutes post-injection, with higher signal in gray matter than white matter, consistent with previous findings.
  • Radiation Dosimetry: The effective dose was calculated to be 4.27 ± 0.57 µSv/MBq. This is approximately 35% of the effective dose of [¹⁸F]3F4AP and comparable to other ¹¹C tracers. The kidneys received the highest organ dose.
  • Safety: No significant changes in vital signs were observed during the PET imaging acquisition, and no delayed adverse reactions were noted in subsequent observations.

Limitations and Future Work

  • Sterility Tests Not Performed: The paper notes that sterility tests were not performed as the radiosynthesis was carried out in a non-sterile environment. This is a standard limitation for radiosynthesis that requires separate sterile formulation steps for clinical use.
  • Limited Sample Size: The biodistribution and dosimetry study was conducted in only two non-human primates. While no sex-based differences were observed, a larger sample size would provide more robust statistical power.
  • Future Work: The authors clearly aim for clinical translation. The successful development of a cGMP-compliant automated synthesis and favorable preclinical dosimetry data lay the groundwork for Investigational New Drug (IND) application and subsequent human clinical trials to evaluate [¹¹C]3MeO4AP for imaging demyelinating diseases. Further studies could also explore its utility in other neurological conditions where potassium channel dysfunction is implicated.

Summary updated

Abstract

The cGMP compliant automated radiosynthesis of [ <sup>11</sup> C]3MeO4AP was developed. The whole-body biodistribution and radiation dosimetry of [ <sup>11</sup> C]3MeO4AP was successfully evaluated in NHPs. [ <sup>11</sup> C]3MeO4AP shows lower average effective dose than [ <sup>18</sup> F]3F4AP and similar average effective dose as other carbon-11 tracers.
BibTeX
@misc{zhou2023radiosynthesis,
  author = {Yu‐Peng Zhou and Moses Q. Wilks and Maëva Dhaynaut and Nicolas J. Guehl and Sung‐Hyun Moon and Georges El Fakhri and Marc D. Normandin and Pedro Brugarolas},
  title = {{Radiosynthesis automation, non-human primate biodistribution and dosimetry of K <sup>+</sup> channel tracer [ <sup>11</sup> C]3MeO4AP}},
  year = {2023},
  doi = {10.1101/2023.03.28.534386},
  howpublished = {bioRxiv},
}

Loading related content...

Figure