Quasi-Static Calibration Method of a High-g Accelerometer

Yan Wang, Jinbiao Fan, Jing Zu, Peng Xu2017

TL;DR

This paper presents a novel quasi-static calibration method for high-g accelerometers, addressing the challenge of resonance during traditional calibration. By analyzing the accelerometer's second-order model and deriving the minimum excitation pulse width, they demonstrate a gas-gun-based system that achieves accurate calibration for accelerometers with resonant frequencies above 20 kHz, with a calibration error of 3%.

What Problem Does This Paper Address?

High-g accelerometers are crucial for measuring extreme accelerations encountered in applications like projectile impacts. Accurate calibration of these sensors is vital for reliable data acquisition. However, traditional calibration methods, especially quasi-static ones, struggle with high-g accelerometers due to their high resonant frequencies. When subjected to short, high-amplitude excitation pulses, these accelerometers can resonate, leading to inaccurate measurements of impact sensitivity. Existing methods often rely on complex filtering or struggle to produce sufficiently long pulses to avoid resonance for lower-frequency high-g sensors. This paper aims to improve quasi-static calibration theory and practice to overcome these resonance-induced errors.

Methods

The core of this work lies in improving the quasi-static calibration theory and implementing a practical testing system.

  • Theoretical Improvement: The authors model the high-g accelerometer as a second-order spring-mass-damper system. By analyzing this model, they derive a relationship between the minimum excitation pulse width, the accelerometer's resonant frequency, and the acceptable calibration error. This relationship is crucial for determining the appropriate pulse width to avoid exciting the accelerometer's resonant modes.
  • Experimental Setup: A quasi-static calibration testing system is developed, utilizing a gas gun to generate high-g acceleration signals. A laser interferometer is employed to accurately measure the excitation acceleration. The system uses a projectile impacting a collision object (where the accelerometer is mounted) to produce the high-g pulse. The pulse width is controlled by adjusting parameters of the collision, such as the contact area and the thickness of a felt pad.
  • Calibration Procedure: The system generates wide-pulse acceleration signals using the gas gun. The laser interferometer provides the reference excitation acceleration signal, while the calibrated accelerometer outputs its response. By comparing these two signals, the impact sensitivity of the accelerometer is determined.

Datasets

Not discussed in the paper. The paper focuses on a single calibration test of a specific piezoelectric high-g accelerometer.

Key Insights and Ideas

  • Resonance is the Enemy: The paper clearly illustrates how short excitation pulses can excite the resonant frequencies of high-g accelerometers, leading to significant errors in impact sensitivity measurements.
  • Pulse Width is Key: The derived relationship between pulse width, resonant frequency, and calibration error is a critical theoretical contribution. It provides a quantitative basis for selecting appropriate excitation pulse durations.
  • Gas Gun as a Controlled Impact Source: The use of a gas gun, combined with careful control of collision parameters (contact area, felt pad thickness), allows for the generation of controlled, wide-pulse high-g acceleration signals.
  • Laser Interferometry for Absolute Measurement: Employing a laser interferometer for measuring the excitation acceleration ensures traceability and accuracy, which is essential for absolute calibration.
  • Wide Pulse = Less Filtering: The advantage of using wide pulses is that it minimizes the need for signal filtering, which can itself introduce errors and alter the peak values of the acceleration signal.

Key Results

  • The developed system can produce excitation acceleration signals with pulse widths greater than 300 µs, and in some cases, up to 1000 µs.
  • The quasi-static calibration method, when applied with appropriate pulse widths, can achieve calibration errors as low as 3% for high-g accelerometers with resonant frequencies above 20 kHz.
  • The paper demonstrates successful calibration of a piezoelectric high-g accelerometer, achieving impact sensitivities within an acceptable error range.
  • Simulations using ANSYS-LS-DYNA helped in understanding how collision parameters influence the excitation pulse width and amplitude, guiding the experimental design.

Limitations and Future Work

  • Specific Accelerometer Type: The calibration was performed on a piezoelectric-type high-g accelerometer. While the principles are generalizable, further validation on other types (e.g., piezoresistive) might be beneficial.
  • Limited Calibration Range: The paper focuses on a specific calibration error of 3% and a resonant frequency above 20 kHz. Extending this to a wider range of errors and lower resonant frequencies would be valuable.
  • Single Test Case: While the calibration data is presented, it's from a single accelerometer. A broader statistical analysis across multiple devices would strengthen the findings.
  • Future Work: The authors suggest that this method can satisfy the demands of engineering tests, implying its readiness for practical application. Further research could explore optimizing the gas gun system for even finer control over pulse characteristics and extending the calibration capabilities to even higher g-levels or different accelerometer technologies.

Summary updated

BibTeX
@misc{wang2017quasi,
  author = {Yan Wang and Jinbiao Fan and Jing Zu and Peng Xu},
  title = {{Quasi-Static Calibration Method of a High-g Accelerometer}},
  year = {2017},
  doi = {10.3390/s17020409},
}