Quantitative Measurement of Tau Burden in a Dual-Time-Window Dynamic PET Imaging Protocol with [18F]MK6240

Ye Xia, Maëva Dhaynaut, Yanis Chemli, Cristina Lois, Bernard Hanseeuw, Emma G. Thibault, Colin Groot, Rik Ossenkoppele, Keith A. Johnson, Georges El Fakhri, Marc D. Normandin, Nicolas J. GuehlJournal of Nuclear Medicine 2025 · DOI

This study, published in The Journal of Nuclear Medicine, provides a practical solution to one of the most significant hurdles in neuroimaging research: the grueling length of dynamic PET scans. By validating a "split-scan" protocol, the authors show we can maintain gold-standard data quality while giving patients a much-needed break.

TL;DR

Researchers validated a dual-time-window (DTW) protocol for $[^{18}F]MK6240$ PET imaging that allows for a 60-minute break during a 120-minute procedure. This approach yields tau burden measurements nearly identical to continuous 2-hour scans while improving patient comfort and scanner throughput.

What Problem Does This Paper Address?

Quantifying tau protein aggregates is essential for tracking Alzheimer’s disease progression. The "gold standard" for this is dynamic PET imaging, which involves scanning the patient continuously for 120 minutes to calculate the Distribution Volume Ratio (DVR). This method is highly accurate and provides "free" data on cerebral perfusion (blood flow) from the early phase of the scan.

However, asking elderly patients—many with cognitive impairment—to lie perfectly still for two hours is a tall order. The common alternative is a "static" scan (SUVR) taken late in the process, which is shorter but less accurate, prone to bias from blood flow changes, and misses the early-phase perfusion data. The challenge was to find a middle ground: a protocol that is as accurate as a full dynamic scan but as tolerable as a static one.

Methods

The researchers didn't just guess at a shorter protocol; they used a retrospective "leave-out" analysis.

  1. Simulation: They took existing 120-minute continuous PET datasets and "deleted" the middle sections to simulate breaks of 30, 40, 50, 60, and 70 minutes.
  2. Interpolation: To calculate the DVR, you need a continuous curve of radioactivity over time. The team tested two ways to "bridge the gap" created by the break: simple linear interpolation and a more complex biexponential model fit.
  3. Validation: They compared the results from these simulated "break" protocols against the original full 120-minute "ground truth" data. They looked at both regional brain averages and voxel-by-voxel parametric maps.

Datasets

The study utilized a dataset of 25 human participants:

  • 13 cognitively normal (CN) individuals.
  • 10 with mild cognitive impairment (MCI).
  • 2 with Alzheimer’s disease (AD).
  • Reference Data: All participants had full 120-minute $[^{18}F]MK6240$ dynamic PET scans, structural MRIs, and $[^{11}C]PiB$ PET scans to confirm amyloid status.
  • Reliability: Two CN subjects underwent "test-retest" scans to ensure the DTW method was stable over time.

Key Insights and Ideas

  • Biexponential is King: While linear interpolation worked for short 30-minute breaks, it failed as breaks got longer. The biexponential model, which better reflects the underlying biological "two-tissue" movement of the tracer, remained highly accurate even with long gaps.
  • The 60-Minute "Sweet Spot": A 60-minute break emerged as the optimal compromise. It is long enough to allow a clinic to scan a different patient in the interim (increasing efficiency) but short enough that the mathematical model can still perfectly predict the missing data.
  • No Data Left Behind: Unlike static scans, this DTW protocol preserves the early-phase data (0–20 mins), allowing researchers to measure cerebral perfusion alongside tau burden in a single session.

Key Results

  • High Accuracy: For the 60-minute break protocol using biexponential fitting, the mean difference in DVR compared to the full scan was less than 1%.
  • Low Variability: The standard deviation of these differences was less than 4%, and there was no statistically significant difference between the DTW and full acquisition results ($P > 0.05$).
  • Visual Consistency: Parametric maps (the actual images of the brain) generated from the DTW protocol were visually and quantitatively indistinguishable from those generated from the full 2-hour scan.
  • Reliability: Test-retest variability for the DTW protocol was comparable to that of the full dynamic scan.

Limitations and Future Work

  • Post-hoc Simulation: The study used data from patients who stayed on the bed. In a real DTW scenario, the patient leaves and returns. While the authors suggest CT scans can handle the realignment, this "real-world" movement wasn't physically tested here.
  • Sample Size: With only 25 subjects and only 2 AD patients, the model needs further testing in cohorts with very high tau burdens to ensure the biexponential fit holds up when tracer kinetics are extremely slow.
  • Physiological Shifts: There is a theoretical risk that a patient’s physiology (like blood pressure or heart rate) could change during a 60-minute break, potentially affecting the late-phase data in ways a continuous scan wouldn't see.

Overall, this paper provides a robust framework for making high-end Alzheimer's research more patient-friendly and clinically viable without sacrificing the precision of the data.

Summary updated

Abstract

This study aimed to test and validate a dual-time-window (DTW) protocol for 6-(fluoro-<sup>18</sup>F)-3-(<sup>1</sup>H-pyrrolo[2,3-c]pyridin-1-yl)isoquinolin-5-amine ([<sup>18</sup>F]MK6240) dynamic PET imaging in experimental datasets acquired in human subjects. <b>Methods:</b> DTW protocols were tested and validated in datasets previously collected in 25 participants: 13 were cognitively normal, 10 had mild cognitive impairment, and 2 had Alzheimer disease. Participants underwent full 120-min [<sup>18</sup>F]MK6240 dynamic PET scans as well as structural MRI. Intermediary 3-dimensional volumes were removed from the acquired dynamic PET images to emulate DTW acquisitions consisting of an early phase and a late phase. Five break durations (30, 40, 50, 60, and 70 min) were investigated to determine the optimal break for 2 study durations (120 and 110 min). Regional brain time-activity curves were extracted using atlases available in the Montreal Neurologic Institute template space and using the FreeSurfer parcellation. Interpolation strategies were tested to recover the missing time points. Distribution volume ratio (DVR) estimates obtained from the DTW time-activity curves were compared with those obtained from the full time-activity curves as reference. Parametric maps were generated for the selected protocol and evaluated. <b>Results:</b> The correlation and agreement between DVR values obtained from the DTW method and the full time-activity curves were overall very good. The DTW protocol with a 60-min break using a biexponential model fit as the interpolation method provided the best compromise between practicality and quantitative accuracy. The mean differences between this DTW and the full acquisition, averaged across brain regions and all subjects, were less than 1% with a corresponding SD of less than 4%, and DVR estimates were not statistically different from those obtained from the full acquisition (<i>P</i> > 0.05). DVR parametric images were visually and quantitatively consistent with those obtained from the full acquisition. <b>Conclusion:</b> This study presents strong support for the use of a DTW protocol with [<sup>18</sup>F]MK6240. Such a protocol would be well suited to allow for both quantification of tau and derivation of an index of cerebral perfusion while reducing patient discomfort and increasing scanning efficiency in comparison to a full dynamic acquisition.