Quantitative Infrared Spectroscopy Of Glucose In Blood Using Partial Least-Squares Analyses
Abstract
The concentration of glucose in drawn samples of human blood has been determined using attenuated total reflectance (ATR) Fourier transform infrared (FT-IR) spectroscopy and partial least-squares (PLS) multivariate calibration. A twelve sample calibration set over the physiological glucose range of 50-400 mg/deciliter (dl) resulted in an average error of 5.24 mg/dl. These results were obtained using cross validated PLS calibration for all data in the frequency range of 950-1200 cm-1. These results are a dramatic improvement relative to those obtained by previous studies of this system using univariate peak height analyses.
BibTeX
@inproceedings{ward1989quantitative,
author = {Kenneth J. Ward and David M. Haaland and Matthew R. Robinson and R. Philip Eaton},
title = {{Quantitative Infrared Spectroscopy Of Glucose In Blood Using Partial Least-Squares Analyses}},
booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
year = {1989},
doi = {10.1117/12.969626},
}Ask Althea to put this paper to the test.