Quantitative determination of gadolinium and iodine contrast agents in dual-energy computed tomography via a dual-energy iterative reconstruction algorithm: a simulation study on multi-contrast imaging

Maria Magnusson, Michael Sandborg, Åsa Carlsson Tedgren, Alexandr MalušekRadiation 2025

TL;DR

This paper presents a simulation study demonstrating that a dual-energy iterative reconstruction algorithm (DIRA) can accurately quantify both iodine and gadolinium contrast agents simultaneously in dual-energy CT (DECT) images. This advancement is crucial for improving diagnostic accuracy and streamlining clinical workflows by enabling better differentiation of contrast agent uptake and distribution.

What Problem Does This Paper Address?

Current X-ray imaging primarily relies on iodine-based contrast agents. However, there's a growing interest in using other contrast agents, like gadolinium (commonly used in MRI), for X-ray imaging to offer complementary information. The challenge lies in accurately quantifying multiple contrast agents simultaneously within the same scan, especially when they have different K-edge energies. Traditional CT reconstruction methods struggle with beam-hardening artifacts, which are exacerbated by the presence of multiple materials with distinct attenuation properties. This paper tackles the problem of achieving precise, simultaneous quantification of iodine and gadolinium contrast agents in DECT, overcoming the limitations of existing methods.

Methods

The researchers employed a simulation-based approach using a custom-built DECT imaging chain. The core of their method is the Dual-Energy Iterative Reconstruction Algorithm (DIRA). DIRA is an iterative algorithm that works by simultaneously processing both projection data (raw X-ray measurements) and image data. This hybrid approach allows it to more effectively suppress beam-hardening artifacts compared to purely image-based or projection-based methods.

Key aspects of their DIRA implementation include:

  • Region-Specific Two-Material Decomposition: Instead of assuming a single material basis for the entire image, DIRA was configured to use different material pairs (e.g., water-bone, water-iodine, water-gadolinium) for different segmented regions within the phantom. This is crucial because gadolinium's K-edge energy requires a different basis than iodine for accurate representation.
  • Iterative Refinement: The algorithm iteratively refines the reconstructed images by comparing simulated projections with measured projections and using the residuals to correct the image. This process progressively reduces artifacts and improves quantitative accuracy.
  • Simulation Framework: They used Siemens' DRASIM code to generate simulated X-ray projections at two energy levels (80 kV and Sn140 kV) to mimic a DECT acquisition.

Datasets

The study utilized a simulated phantom for evaluation, not real-world datasets. This phantom was a cylindrical lipid phantom containing five internal inserts filled with:

  • Water
  • Iodine solutions (25 mg/mL and 50 mg/mL)
  • Gadolinium solutions (25 mg/mL and 50 mg/mL)

This simulated setup allowed for precise control over the concentrations and spatial distribution of the contrast agents, serving as a benchmark for the DIRA algorithm's performance.

Key Insights and Ideas

  • Region-Specific Material Decomposition is Key: The study strongly highlights that a "one-size-fits-all" approach to material decomposition in DECT is insufficient when dealing with agents like gadolinium. The K-edge energy of gadolinium necessitates a dedicated basis (water-gadolinium) for accurate quantification, which differs from the optimal basis for iodine (water-iodine).
  • DIRA's Iterative Nature Effectively Combats Artifacts: The iterative nature of DIRA proved highly effective in suppressing beam-hardening artifacts, which are a significant challenge in DECT, especially when dealing with materials that have sharp K-edges. The visual improvement from iteration 0 to iteration 16 is striking.
  • Simultaneous Quantification is Feasible: The research demonstrates that it's possible to accurately determine the mass fractions of both iodine and gadolinium within the same scan, even when they are spatially separated but present in the same phantom. This opens doors for more comprehensive diagnostic imaging.
  • Quantitative Accuracy Achieved Quickly: The algorithm achieved quantitative accuracy close to reference values within approximately six iterations, suggesting a reasonable trade-off between computational time and image quality for clinical applications.

Key Results

  • Accurate LAC Recovery: Reconstructed linear attenuation coefficients (LACs) for iodine and gadolinium solutions converged to within approximately 1-2% of reference values (from the EPDL97 library) after 16 iterations.
  • Precise Mass Fraction Estimation: The reconstructed mass fractions for both iodine and gadolinium solutions were accurately determined. For iodine, deviations were within 1% of the true values by iteration 6. For gadolinium, initial estimates were significantly improved, reaching agreement within 4% at iteration 6 and 1% at iteration 16.
  • Significant Artifact Reduction: Visualizations clearly show the dramatic reduction of beam-hardening artifacts from the initial reconstruction (iteration 0) to the final iteration (iteration 16), resulting in a more uniform and artifact-free image.

Limitations and Future Work

The authors acknowledge several limitations:

  • Simulation-Only Study: The study was entirely based on simulations, and no experimental data was analyzed. Future work should involve validation with real-world phantom and clinical data.
  • Manual Segmentation: The regions containing the materials were manually defined. Automating this segmentation process using threshold-based or machine learning methods is a crucial next step.
  • Two-Material Assumption: The current DIRA implementation assumes a two-material composition per voxel. This assumption breaks down when contrast agents mix or spatially overlap. Future extensions to three-material decomposition (water, iodine, gadolinium) are needed for more complex scenarios.
  • No Quantum Noise: Quantum noise, an inherent aspect of real CT data, was not simulated. While previous studies suggest minor effects on mean values, its inclusion would provide a more complete picture of performance.

Future work should also explore integrating DIRA with photon-counting CT (PCCT) systems, which offer enhanced spectral resolution and could further improve quantitative accuracy and material characterization.

Summary updated

Abstract

Simultaneous quantification of two contrast agents in dual-energy CT (DECT) is promising for clinical diagnostics. This simulation study demonstrates that accurate estimation of iodine and gadolinium concentrations in unmixed aqueous solutions can be achieved using two-material decomposition and beam-hardening suppression with the dual-energy iterative reconstruction algorithm (DIRA). A simulated DECT imaging chain was used for evaluation. The phantom comprised a lipid cylinder containing five inserts filled with water, iodine (25 and 50 mg/mL), and gadolinium (25 and 50 mg/mL) solutions. X-ray projections at 80 and Sn140 kV were generated using Siemens' DRASIM code. DIRA was configured for two-material decomposition with water-bone, water-iodine, and water-gadolinium bases, and cylinder positions were manually defined during reconstruction. After approximately six iterations, accurate estimates of contrast agent mass fractions were obtained. These results confirm the feasibility of simultaneous iodine and gadolinium quantification in DECT and support further development toward a clinical application.