Quantifying cardiac‐induced brain tissue expansion using DENSE

Ayodeji L. Adams, Hugo J. Kuijf, Max A. Viergever, Peter R. Luijten, Jaco J.m. ZwanenburgNMR 2018

TL;DR

This paper introduces a 3D cine-DENSE MRI technique to precisely measure tiny, cardiac-induced volume changes in brain tissue. They demonstrate that this method can distinguish between grey and white matter, offering a promising new way to study small blood vessel health and its changes with age and disease.

What Problem Does This Paper Address?

The brain's volume subtly changes with each heartbeat due to blood flow fluctuations in its microvasculature. These changes, known as volumetric strain, are thought to reflect the health of small blood vessels and the surrounding tissue's mechanical properties. However, accurately measuring this tiny strain across the entire brain over the cardiac cycle has been challenging. Previous MRI techniques either lacked the resolution, were limited to 2D, or suffered from noise, preventing a comprehensive understanding of this phenomenon and its potential as a biomarker for conditions like cerebral small vessel disease (cSVD). This study aims to overcome these limitations by implementing and validating a high-resolution 3D cine-DENSE MRI sequence at both 7 Tesla (T) and 3T.

Methods

The core of this research is the application of Displacement Encoding with Stimulated Echoes (DENSE), a specialized MRI technique. They used a 3D cine-DENSE sequence, meaning it captures data throughout the cardiac cycle (cine) in three dimensions. This sequence was implemented on both a 7T and a 3T MRI scanner. Key technical aspects include:

  • Cardiac Gating: The MRI acquisition is synchronized with the subject's heartbeat (using ECG or pulse oximetry) to capture the dynamic changes.
  • Motion Encoding: DENSE works by applying magnetic field gradients that encode motion into the MRI signal's phase. By using opposing gradients in subsequent acquisitions, they could subtract phase errors and isolate the motion-induced phase shifts.
  • Fat Suppression: Specialized radiofrequency (RF) pulses and gradient adjustments were used to suppress signals from fat tissue, which can otherwise interfere with motion measurements.
  • Variable Flip Angle Strategy: To optimize signal-to-noise ratio (SNR) across different phases of the cardiac cycle, they employed a variable flip angle approach, adapting the RF pulse angle based on the tissue's T1 relaxation time.
  • Volumetric Strain Calculation: The measured displacements were then used to compute volumetric strain by calculating the divergence of the displacement field. This essentially measures the rate of expansion or contraction at each point in the brain tissue.
  • SNR Analysis: A crucial part of the study was to assess the SNR of the DENSE measurements and its impact on the accuracy of the volumetric strain calculations, especially comparing the higher field strength (7T) to the lower one (3T).

Datasets

The study utilized data from six healthy human subjects. No new benchmark datasets were introduced; the focus was on applying and validating the DENSE technique on human participants.

Key Insights and Ideas

  • Brain Tissue Pulsates: The study confirms that brain tissue undergoes measurable, albeit small, volumetric expansion and contraction synchronized with the heartbeat.
  • Grey vs. White Matter Differences: A significant finding is the approximately four-times higher peak volumetric strain observed in grey matter compared to white matter. This highlights distinct physiological responses and suggests that DENSE can differentiate between these tissue types based on their pulsatile behavior.
  • 7T Advantage (SNR): The 7T scanner provided a significantly higher SNR (about three times) compared to 3T. This improved SNR leads to less noise in the displacement maps, making the data appear smoother.
  • SNR Limitation for Voxel-wise Analysis: Despite the improved SNR at 7T, it was still not high enough for reliable voxel-wise analysis of volumetric strain at the resolution used (2.2 mm isotropic). This indicates that further improvements in SNR are needed for detailed, single-voxel measurements.
  • Potential Biomarker for Small Vessel Disease: The ability to quantify these subtle volume changes holds promise for detecting early signs of small vessel disease (cSVD), which affects the brain's microvasculature, potentially even before overt lesions appear.

Key Results

  • Whole Brain Volumetric Strain: Mean peak whole brain volumetric strain was measured at (4.5 ± 1.0) × 10⁻⁴ at 7T, corresponding to a volume expansion of approximately 0.48 ± 0.1 mL. This value aligns with literature estimates for cerebrospinal fluid displacement.
  • Grey vs. White Matter Strain Ratio: The peak volumetric strain ratio of grey to white matter was found to be 4.4 ± 2.8 at 7T. This difference was statistically significant (p < 0.001).
  • SNR Performance: The mean SNR for DENSE measurements was 22.0 ± 7.3 at 7T and 7.0 ± 2.8 at 3T, showing a 7T to 3T SNR ratio of 3.2 ± 0.7.
  • Voxel-wise Uncertainty: The expected voxel-wise uncertainty in volumetric strain was calculated to be 1.4 × 10⁻³ at 7T and 4.4 × 10⁻³ at 3T, highlighting the current limitations for detailed analysis.

Limitations and Future Work

  • SNR for Voxel-wise Analysis: The primary limitation acknowledged is the current SNR, which restricts detailed voxel-wise analysis of volumetric strain. The authors suggest that a tenfold increase in SNR would be needed for such analysis.
  • Artefact Removal: While efforts were made to remove artefacts, some residual issues might affect the strain curve's return to baseline, suggesting imperfect artefact removal.
  • Future Directions:
    • Improving SNR: Strategies to boost SNR include using specialized local surface coils, employing de-noising algorithms, increasing voxel size (with caution regarding partial volume effects), and exploring balanced displacement encoding.
  • Acquisition Strategies: Investigating alternative acquisition schemes like spiral readouts or incorporating compressed sensing and parallel imaging could improve efficiency and SNR.
  • Clinical Applications: The authors emphasize the potential for using this technique to study blood volume pulsations in the aging brain and in diseases like cSVD, suggesting future work in patient populations.
  • Normalization: For clinical applications, normalization to cardiac function might be necessary to accurately evaluate changes in brain tissue volumetric strain.
  • Disentangling Contributions: Further research could aim to disentangle the contributions of small vessel swelling from tissue stiffness and other factors influencing volumetric strain.

Summary updated

BibTeX
@inproceedings{adams2018quantifying,
  author = {Ayodeji L. Adams and Hugo J. Kuijf and Max A. Viergever and Peter R. Luijten and Jaco J.M. Zwanenburg},
  title = {{Quantifying cardiac‐induced brain tissue expansion using DENSE}},
  booktitle = {NMR},
  year = {2018},
  doi = {10.1002/nbm.4050},
}
Figure