Quality and Similarity Indicators for Video-Based Heart Rate and Respiratory Rate Fusion: A Controlled Evaluation

Nhi Nguyen, Le Nguyen, Manuel Lage Cañellas, Miguel Bordallo López, Constantino Álvarez CasadoarXiv 2026

TL;DR

This paper presents a rigorous, controlled evaluation of whether reference-free quality indicators can actually improve the fusion of multiple video-based heart rate (HR) and respiratory rate (RR) estimation methods. The authors find that while quality-guided fusion significantly boosts HR accuracy, it fails for RR, where apparent gains often merely reflect a "shrinkage" effect toward typical physiological rates rather than genuine signal tracking.

What Problem Does This Paper Address?

Estimating vital signs like HR and RR from video is notoriously difficult because signals are weak and easily drowned out by motion, lighting changes, or compression artifacts. Since no single algorithm performs best across all conditions, researchers often "fuse" multiple methods, weighting them by reference-free Signal Quality Indicators (SQIs).

However, the field has a validation problem. Many studies report fusion gains without ruling out simpler explanations:

  1. Simple Averaging: Does the gain come from the quality indicators, or just from reducing variance by averaging multiple noisy estimates?
  2. Shrinkage toward the Mean: If estimates are poor, any method that pulls them toward a "typical" rate (e.g., 75 bpm for HR) will lower the error.
  3. The Oracle Trap: Picking the best estimate per window from a large pool will always look good by chance.

The authors set out to determine if these indicators provide genuine, window-specific reliability information that outperforms these simple baselines.

Methods

The researchers developed a comprehensive framework to evaluate 7 HR and 10 RR extraction methods (ranging from classical color-projection techniques like CHROM to deep learning models like PhysFormer).

Figure 1: Quality and similarity framework. Candidate HR and RR signals are extracted by classical, motion-based and deep learning methods, filtered over the full recording, converted to rates in 10-s windows by MUSIC, scored by quality indicators, inter-method similarity or a learned ridge score, and fused or filtered. Steps in the dashed box are fixed on the five-subject Trainset before any test set is evaluated. Fusion is evaluated against a common reference and against averaging, a constant predictor and a null oracle.

Key methodological innovations include:

  • Indicator Pool: They evaluated 11 intrinsic quality indicators (e.g., SNR, skewness, template matching) and 8 extrinsic similarity measures (how much different methods agree with each other).
  • Strict Controls: They compared fusion against equal averaging, a constant "typical-rate" predictor, and a Null Oracle (a permutation-based control that measures the advantage of picking the best estimate by pure chance).
  • Fixed-in-Advance Protocol: To prevent over-fitting, all choices—from filter chains to indicator directions—were "locked in" using a small 5-subject training set before any testing on the main datasets occurred.

Datasets

The study utilized four diverse public datasets comprising 75 test subjects:

  • OMuSense-23: Used for model training and indicator selection.
  • COHFACE, MAHNOB-HCI, and StressID: Used exclusively for independent testing. These datasets cover various conditions, including resting states, emotional elicitation, and stress-inducing tasks.

Key Insights and Ideas

  • HR vs. RR Disparity: Quality indicators are highly effective for HR but fail for RR. This is likely due to "cycle count": a 10-second window contains 10–15 heartbeats (providing enough data for a robust quality estimate) but only 1–3 breaths, making RR quality much harder to assess.
  • The "Typical Rate" Baseline: For RR, a simple constant predictor (always guessing the median rate) was more accurate than equal averaging. This reveals that many "successful" RR fusion results in literature might just be accidentally benefiting from pulling estimates toward the center of the physiological band.
  • Similarity isn't enough: Weighting methods by how much they agree with each other (consensus) did not consistently beat simple averaging. If all methods are wrong in the same way (e.g., tracking head motion instead of pulse), agreement becomes a false signal of quality.
  • Partial Correlation is Crucial: The authors used partial correlations to show that for RR, indicators were mostly just "restating" the estimated rate. Once the predicted rate was controlled for, the indicators' link to actual error largely vanished.

Key Results

  • Heart Rate Success: Quality-weighted fusion reduced Mean Absolute Error (MAE) by 10.9% relative to simple averaging and by 15.6% relative to the best pre-selected single method. It performed as well as the "best method in hindsight."
  • Respiratory Rate Failure: In 10-second windows, no indicator improved fusion. In fact, "sharpening" the weights based on quality indicators actually increased error by 2.6%.
  • Filtering Gains: Using quality scores to simply discard the worst 30% of windows (rather than fusing) improved HR accuracy by up to 25% compared to random removal.
Figure 2: Single-method MAE (top rows) and Pearson 𝑟(bottom rows) for each method and rate estimator. The Trainset-chosen estimator (MUSIC, bold) is used for everything that follows; red labels mark the Trainset DN1 pair (HR: POS + PF; RR: Mot-S + PF).

Limitations and Future Work

The authors acknowledge that 10-second windows may simply be too short for reliable RR quality assessment. While they tested 30-second windows and saw improved base estimates, quality-guided fusion still didn't beat simple averaging.

Future work might explore:

  • Breath-by-breath morphology: Moving away from spectral estimates toward temporal analysis of individual breaths.
  • Causal Filtering: This study used offline filtering; real-time applications will need to validate these findings with causal, low-latency pipelines.
  • Task-Specific Models: Deep models often struggled with "out-of-distribution" data (e.g., StressID), suggesting a need for better generalization or domain adaptation in the underlying extractors.

Summary updated

Abstract

Camera-based heart rate (HR) and respiratory rate (RR) estimates vary in reliability from window to window, and a common remedy is to fuse several extraction methods with weights from reference-free signal quality indicators. However, few studies test whether these indicators identify the most accurate method in each window or outperform simpler fusion. We evaluate 11 quality indicators and 8 inter-method similarity measures for fusing 7 HR and 10 RR extractors on four public datasets (75 test subjects). Filters, indicator directions, weighting parameters and a learned ridge baseline are selected on a separate five-subject set before testing. Fusion is compared with equal averaging, a constant typical-rate predictor, the best single method selected in advance and in hindsight, and a per-window oracle with a permutation-based null, using subject-level bootstrap confidence intervals. For HR, quality-weighted fusion reduced mean absolute error by 10.9% [5.2, 18.4] relative to averaging and by 15.6% [8.1, 23.1] relative to the method selected in advance, and was statistically indistinguishable from the best method selected in hindsight. Weighting methods by their agreement did not improve on averaging. For RR, the indicators could not identify the more accurate estimates: sharpened quality weighting increased error (-2.6% [-4.8, -0.8]), fused estimates were uncorrelated with the reference (r <= 0.22), and averaging was less accurate than the typical-rate predictor (-3.1% [-6.1, -0.4]). With a second respiratory reference, apparent RR gains reflected estimates moving toward its typical rate. Quality-guided fusion is worthwhile for video HR; for RR, 30-s windows improved estimates but not fusion.
BibTeX
@misc{nguyen2026quality,
  author = {Nhi Nguyen and Le Nguyen and Manuel Lage Cañellas and Miguel Bordallo López and Constantino Álvarez Casado},
  title = {{Quality and Similarity Indicators for Video-Based Heart Rate and Respiratory Rate Fusion: A Controlled Evaluation}},
  year = {2026},
  eprint = {2611.21419},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/2611.21419},
}
Figure