Prostate Brachytherapy Seed Reconstruction Using C-Arm Rotation Measurement and Motion Compensation
Lists
Only you see this
Abstract
During prostate brachytherapy, C-arm flouroscopy images are used for a qualitative assessment of the procedure. Three dimensional reconstruction of the implanted seeds can be used for intraoperative dosimetry and quantitative assessment. Accurate C-arm pose estimation is necessary for 3D localization of the seeds. We propose to measure the C-arm rotation angles and computationally compensate the inevitable C-arm translational motion to estimate the pose. We compensate the oscillation, sagging and wheel motion of the C-arm using a three-level optimization algorithm, without which the reconstruction can fail. We validated our approach on simulated and 10 data sets from 5 patients and gained on average 99.1% success rate, 0.33 mm projection error and computation time of less than one minute per patient, which are clinically excellent results.
BibTeX
@inproceedings{degan2010prostate,
author = {Ehsan Degan and Junghoon Lee and Mehdi Moradi and Xu Wen and Gabor Fichtinger and Septimiu E. Salcudean},
title = {{Prostate Brachytherapy Seed Reconstruction Using C-Arm Rotation Measurement and Motion Compensation}},
booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
year = {2010},
url = {https://openreview.net/forum?id=YppdV-N1g87},