Precision pH Sensor Based on WO3 Nanofiber-Polymer Composites and Differential Amplification
Lists
Only you see this
Abstract
We report a new type of potentiometric pH sensor with sensitivity exceeding the theoretical Nernstian behavior (-59.1 mV/pH). For the pH-sensitive electrode, 1D tungsten oxide (WO<sub>3</sub>) nanofibers (NFs) were prepared to obtain large surface area and high porosity. These NFs were then stabilized in a reactive porous chloromethylated triptycene poly(ether sulfone) (Cl-TPES) binder, to facilitate proton diffusion into the polymer membrane. The measurements were performed with a differential amplifier using matched MOSFETs and providing a 10-fold amplified signal over a simple potentiometric determination. A high pH sensitivity of -377.5 mV/pH and a linearity of 0.9847 were achieved over the pH range of 6.90-8.94. Improved signal-to-noise ratios with large EMF signal changes of 175 mV were obtained in artificial seawater ranging pH 8.07-7.64 (ΔpH = 0.43), which demonstrates a practical application for pH monitoring in ocean environments.
BibTeX
@inproceedings{choi2019precision,
author = {Seon‐Jin Choi and Suchol Savagatrup and Yoonseob Kim and Jeffrey H. Lang and Timothy M. Swager},
title = {{Precision pH Sensor Based on WO<sub>3</sub> Nanofiber-Polymer Composites and Differential Amplification}},
booktitle = {ACS},
year = {2019},
doi = {10.1021/acssensors.9b01579},
}Loading related content...