Point-spread functions for backscattered imaging in the scanning electron microscope

Philipp Hennig, Winfried DenkJournal of Applied Physics 2007 · DOI

Abstract

One knows the imaging system's properties are central to the correct interpretation of any image. In a scanning electron microscope regions of different composition generally interact in a highly nonlinear way during signal generation. Using Monte Carlo simulations we found that in resin-embedded, heavy metal-stained biological specimens staining is sufficiently dilute to allow an approximately linear treatment. We then mapped point-spread functions for backscattered-electron contrast, for primary energies of 3 and 7 keV and for different detector specifications. The point-spread functions are surprisingly well confined (both laterally and in depth) compared even to the distribution of only those scattered electrons that leave the sample again.
BibTeX
@article{hennig2007point,
  author = {Philipp Hennig and Winfried Denk},
  title = {{Point-spread functions for backscattered imaging in the scanning electron microscope}},
  journal = {Journal of Applied Physics},
  year = {2007},
  doi = {10.1063/1.2817591},
}

Loading related content...

Figure