Photopolymerization 3D-Printed Dual-Modal Flexible Sensor for Glucose and pH Monitoring

Lin Shao, Yu Li, Zhenyao Yang, Qiuzheng Li, Baojun Pang, Feng Yin, Jiao Fu, Guangmeng Ma, Yu LongSensors 2025 · DOI

TL;DR

This paper introduces a novel 3D-printed, flexible sensor capable of simultaneously monitoring glucose and pH levels in physiological fluids. Leveraging a dual-modal approach with microneedle electrodes and advanced photopolymerization, this sensor offers a more comprehensive and less invasive way to track key health indicators compared to existing single-parameter devices.

What Problem Does This Paper Address?

The demand for continuous, real-time monitoring of physiological parameters like glucose and pH is rapidly growing, especially in healthcare and wearable technology. Current flexible sensors often focus on detecting only one analyte at a time. This limitation makes it challenging to get a holistic picture of a patient's metabolic status or tissue environment, as glucose and pH are intrinsically linked and their simultaneous changes can provide crucial diagnostic information. Furthermore, existing multi-analyte sensors can suffer from signal interference, complex fabrication, and high costs. This paper aims to bridge this gap by developing a flexible, high-precision, and cost-effective sensor that can accurately and simultaneously measure both glucose and pH.

Methods

The researchers employed a sophisticated 3D printing technique called Digital Light Processing (DLP) to fabricate a flexible sensor. The core of the sensor is a custom-developed ionic gel substrate, synthesized from polyurethane acrylate (PUA) and acryloylmorpholine (ACMO), which provides mechanical robustness and high ionic conductivity.

Here's a breakdown of the key methodological aspects:

  • 3D Printing with DLP: This technique allowed for the precise fabrication of microneedle arrays and microgrooves with micron-level accuracy, enabling complex structures that are difficult to achieve with traditional methods.
  • Dual-Modal Sensing Mechanism:
    • Glucose Detection: This relies on the enzymatic activity of glucose oxidase (GOD). GOD catalyzes the oxidation of glucose, producing hydrogen peroxide (H₂O₂). The sensor measures the current generated by the electrochemical oxidation of H₂O₂.
  • pH Detection: This utilizes the conductive properties of polyaniline (PANI). PANI's electrical resistance changes based on its protonation state, which is directly influenced by the hydrogen ion concentration (pH).
  • Microneedle Electrode Design: The microneedle array design allows for deeper penetration into the interstitial fluid, improving detection accuracy and speed while minimizing discomfort.
  • Signal Decoupling Strategy: A critical aspect was achieving minimal interference between the glucose and pH signals. This was accomplished by operating the sensor at an optimal potential of 0.2 V, where the glucose oxidation current dominates and the PANI's polarization effect is minimized. Additionally, a Nafion membrane was used to selectively allow protons to pass while limiting H₂O₂ interaction with PANI.
  • Material Synthesis: The development of a stable and conductive ionic gel was crucial. This involved optimizing the ratios of PEG, IPDI, PCL, HEA, ACMO, and the ionic liquid [EMIM][TFSI] to achieve desired mechanical and electrical properties.

Datasets

Not explicitly discussed in the paper as distinct datasets for training or benchmarking. The evaluation was performed using prepared glucose solutions and pH buffers, including artificial interstitial fluid (SBF), to mimic physiological conditions.

Key Insights and Ideas

  • Orthogonality Achieved: The sensor demonstrates excellent orthogonality between glucose and pH measurements, meaning that changes in one analyte have minimal impact on the detection of the other, especially within the physiological pH range.
  • Optimal Operating Potential: Identifying the 0.2 V operating potential was key to decoupling the electrochemical signals from glucose oxidation and PANI's pH-dependent resistance.
  • Ionic Gel as a Versatile Substrate: The developed ionic gel material proved to be a robust and conductive platform, offering tunable mechanical properties and enabling high-resolution 3D printing.
  • Microneedle Advantage: The microneedle design facilitates direct access to interstitial fluid, enhancing sensing performance and user comfort.
  • Synergistic Dual-Modal Design: By integrating two distinct sensing mechanisms onto a single platform, the sensor provides a more comprehensive physiological assessment than single-analyte devices.

Key Results

  • Glucose Sensing: The sensor exhibited a linear response to glucose in artificial interstitial fluid within the range of 0.5–2.5 g·L⁻¹ with a sensitivity of 0.288 µA·mM⁻¹·cm⁻².
  • pH Sensing: The sensor showed a piecewise-linear sensitivity for pH, with 0.155 Ω/pH·cm² for pH > 7 and 0.135 Ω/pH·cm² for pH < 7. The effective pH sensing range was found to be 4–8.
  • Low Cross-Interference: In the physiological range (0.5–2.5 g·L⁻¹ glucose and pH 6.86–7.6), the cross-interference between the two parameters was less than ±20%.
  • Mechanical Properties: The optimized ionic gel exhibited good mechanical strength (0.56 MPa), a breaking strain of 317%, and ionic conductivity of 0.12 mS/cm.
  • Real-World Simulation: In SBF, the sensor demonstrated good accuracy and linearity for glucose concentration measurement over 2 hours, comparable to commercial blood glucose meters. pH measurements remained stable within the physiological range.

Limitations and Future Work

  • Real Penetration Performance: While finite element analysis provided insights into microneedle mechanics, the actual penetration performance on skin-like materials remains unvalidated.
  • Morphological Changes: The study did not experimentally characterize the morphological changes of microneedles before and after use, nor the impact of structural damage on sensing performance.
  • Measurement Range and Time Constraints: The authors acknowledge limitations in measurement accuracy, range, and time constraints.
  • Enzyme Stability: The enzyme reaction in SBF led to data distortion after approximately 2 hours, suggesting potential limitations in long-term enzyme stability in complex biological fluids.
  • Future Directions: The authors propose future work focusing on:
  • Conducting penetration tests on skin-mimicking phantoms.
  • Using SEM to characterize microneedle morphology evolution.
  • Establishing correlations between structural integrity and sensor performance.
  • Exploring multi-functional integration, self-energy supply, and biocompatibility for wearable health detection.

Summary updated

Abstract

Currently, flexible sensors based on electrochemical principles are predominantly limited to single-parameter detection, making it challenging to meet the demand for synchronous monitoring of multiple analytes in complex physiological environments. This study presents a 3D-printed flexible sensor for synchronous glucose/pH detection. Glucose was quantified via H<sub>2</sub>O<sub>2</sub> oxidation current (GOD-catalyzed reaction), while pH was measured through polyaniline (PANI) resistance changes. The ionogel-based microneedle electrode ensures mechanical robustness. At 0.2 V, optimal signal decoupling was achieved: glucose oxidation current dominates, while PANI's polarization effect is minimized. Neutral pH minimally affected glucose oxidase (GOD) activity, and low glucose concentrations induced negligible pH interference, ensuring orthogonality. In artificial interstitial fluid, the sensor showed glucose: linear response (0.5-2.5 g·L<sup>-1</sup>, 0.288 μA·mM<sup>-1</sup>·cm<sup>-2</sup>); pH: piecewise-linear sensitivity (0.155 Ω/pH·cm<sup>2</sup> for pH > 7; 0.135 Ω/pH·cm<sup>2</sup> for pH < 7). The design enables real-time multiparameter monitoring with high selectivity, addressing current limitations in flexible electrochemical sensors.
BibTeX
@inproceedings{shao2025photopolymerization,
  author = {Lin Shao and Yu Li and Zhenyao Yang and Qiuzheng Li and Baojun Pang and Feng Yin and Jiao Fu and Guangmeng Ma and Yu Long},
  title = {{Photopolymerization 3D-Printed Dual-Modal Flexible Sensor for Glucose and pH Monitoring}},
  booktitle = {Sensors},
  year = {2025},
  doi = {10.3390/s25175358},
}
Figure