Phantom-based quantification of the spectral accuracy of lower extremity vascular imaging using a second-generation dual-layer spectral CT at 100 kVp and 120kVp

Xinyu Gao, Chao Zhang, Yun Wang, Xiaomei Lu, Shenghui Yu, Wenjie Zheng, Zehong Lin, Yining Wang, Huadan Xue, Daming Zhang, Zhengyu JinBMC Medical Imaging 2025 · DOI

Introduction

Peripheral artery disease (PAD) is a common condition often diagnosed using Computed Tomography Angiography (CTA). However, standard CTA protocols for the lower limbs require extensive scanning ranges, leading to high cumulative doses of ionizing radiation. Furthermore, achieving adequate vascular enhancement typically necessitates high concentrations of iodine-based contrast media (CM). This poses a significant risk for PAD patients, who are frequently elderly and may have impaired renal function, increasing the likelihood of contrast-induced acute kidney injury.

This study evaluates the potential of second-generation dual-layer spectral CT (DSCT) to mitigate these risks. By leveraging spectral data, specifically Virtual Monoenergetic Images (VMIs), researchers aimed to determine if diagnostic-quality imaging could be maintained at reduced tube voltages and significantly lower iodine concentrations.

Methods

The study utilized a specialized spectral CT phantom containing tissue-equivalent inserts to simulate lower-extremity vasculature. The experimental design systematically varied vessel diameters (2.5, 4, and 5 mm), iodine concentrations (4, 8, 12, and 16 mgI/mL), tube voltages (100 and 120 kVp), and radiation doses (CTDIvol of 2.5, 5, and 10 mGy).

Representative axial CT images of the spectral phantom across various reconstruction and acquisition parameters.

Scans were performed on a Philips Spectral CT 7500. The resulting spectral data allowed for the reconstruction of conventional CT images and VMIs at 40, 70, and 100 keV. Objective image quality was assessed via CT attenuation (HU), standard deviation (noise), Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR). Subjective quality was independently graded by two radiologists using a 5-point Likert scale.

Results

The investigation revealed that second-generation DSCT maintains consistent image quality across varying acquisition parameters. Statistical analysis showed no significant differences in subjective scores or objective metrics (CT value, SD, SNR) when comparing 100 kVp to 120 kVp, or when reducing the radiation dose from 10 mGy to as low as 2.5 mGy.

Quantitative comparison of CT value, SD, CNR, and SNR across different tube voltages and radiation dose levels.

A critical finding was the "iodine amplification" effect of low-energy VMIs. At 40 keV, which is close to the k-edge of iodine (33.2 keV), the attenuation of the contrast agent is significantly enhanced. This allowed an ultra-low iodine concentration of 4 mgI/mL to produce image quality comparable to conventional scans using 12–16 mgI/mL. This effect was particularly pronounced in small vessels; for 2.5 mm simulated vessels at 2.5 mGy, 40 keV VMIs achieved a mean CT value of 626.2 HU and an SNR of 147.9, vastly outperforming conventional CT (269.5 HU and SNR of 47.4).

Discussion and Conclusion

The results demonstrate that 100 kVp can serve as a reliable standard for spectral CT, inherently reducing radiation exposure compared to the 120–140 kVp required by earlier generations. Furthermore, the ability to achieve high-quality vascular imaging at 2.5 mGy and ultra-low iodine concentrations suggests a safer clinical pathway for patients with renal insufficiency.

While these phantom results are promising, limitations include the static nature of the model, which does not account for patient motion or complex pathologies like severe calcification. Future research should focus on anthropomorphic models and clinical trials to validate these low-dose, low-contrast protocols in human subjects.