Abstract
First investigations regarding dosimetric properties of the hybrid, pixelated, photon-counting Dosepix detector in the direct beam of a pulsed photon field (RQR8) for the personal dose equivalent $H_p(10)$ are presented. The influence quantities such as pulse duration and dose rate were varied, and their responses were compared to the legal limits provided in PTB-A 23.2. The variation of pulse duration at a nearly constant dose rate of about 3.7 Sv/h shows a flat response around 1.0 from 3.6 s down to 2 ms. A response close to 1.0 is achieved for dose rates from 0.07 Sv/h to 35 Sv/h for both pixel sizes. Above this dose rate, the large pixels (220 μm edge length) are below the lower limit. The small pixels (55 μm edge length) stay within limits up to 704 Sv/h. The count rate linearity is compared to previous results, confirming the saturating count rate for high dose rates.
Introduction: The Challenge of Pulsed X-Ray Fields
In medical environments like interventional radiology, medical staff are frequently exposed to pulsed X-ray fields. To ensure safety, personnel wear Active Personal Dosemeters (APDs) that provide real-time monitoring. However, most commercial APDs, typically based on silicon diodes or Geiger-Müller tubes, suffer from a critical weakness: dead-time.
When radiation pulses are extremely intense and short, these detectors become overwhelmed. They continue processing the initial photons while subsequent ones arrive, leading to a significant underestimation of the actual dose. This creates a dangerous "false sense of security." This study evaluates whether the Dosepix detector, with its unique hybrid architecture, can overcome these limitations and meet the strict legal metrology standards defined by PTB-A 23.2.
The Dosepix Architecture
The Dosepix is a hybrid pixel detector where a segmented semiconductor sensor is "bump-bonded" to an Application-Specific Integrated Circuit (ASIC). This design allows for sophisticated signal processing at the individual pixel level.
Key technical features include:
- Dual Pixel Sizes: The chip incorporates both "large" pixels (220 µm) and "small" pixels (55 µm). The smaller pixels have a lower detection volume, making them significantly less susceptible to "pile-up" (multiple photons hitting the same pixel simultaneously) during high-intensity pulses.
- Energy Binning: Instead of simple hit counting, the detector sorts events into 16 energy bins, enabling more accurate calculation of the personal dose equivalent $H_p(10)$.
- Rolling-Shutter Readout: This mechanism allows the detector to operate virtually dead-time free, as parts of the matrix can be read out while others continue processing signals.
Performance and the "Small Pixel" Advantage
Experimental tests were conducted at the PTB (Germany’s national metrology institute) using a specialized pulsed X-ray facility (GESA) with a standard medical X-ray quality (RQR8).
One of the most significant findings is the detector's independence from pulse duration. Because of the dead-time-free architecture, the response remained flat across durations ranging from 2 milliseconds to 3.6 seconds. This is vital for medical applications where X-ray pulses often occur in the millisecond range.
Regarding dose rate performance, the dual-pixel design acts as a range-extender. While the large pixels provide high sensitivity for low-background environments, the small pixels serve as a high-range safety valve.
As shown in the figure above, the large pixels remain accurate up to 35 Sv/h. Beyond this, they begin to saturate. However, the small pixels remain within legal limits up to 704 Sv/h. This outperforms most commercial APDs, which typically begin to fail or require complex correction factors at dose rates as low as 1–4 Sv/h, by two orders of magnitude.
Conclusion and Future Outlook
The Dosepix detector demonstrates exceptional potential for personal dosimetry in pulsed fields. The discrepancy between large and small pixel readings can even serve as an internal self-validation mechanism; if large pixels saturate, the system can automatically switch to small-pixel data to maintain accuracy.
Future work will focus on testing the detector at higher energies, such as those found in radiotherapy Linacs (MeV range), and addressing the engineering challenges of shrinking this "dosemeter demonstrator" into a wearable, battery-powered commercial device.
Summary updated