Personal Dosimetry in Continuous Photon Radiation Fields With the Dosepix Detector

Dennis Haag, Sebastian Schmidt, P. Hufschmidt, Franziska Eberle, Thilo Michel, G. Anton, Oliver Hupe, Jürgen Röth, Christian Fuhg, Hayo Zutz, R. Behrens, M. Campbell, X. Llopart, R. Ballabriga, L. Tlustos, W. C. WongIEEE Transactions on Nuclear Science 2021 · DOI

Abstract

First measurements characterizing dosimetric properties of a dosimetry system designed for the purpose of active personal dosimetry for photons with mean energies from 12.4 to 1250 keV according to Physikalisch-Technische Bundesanstalt (PTB) requirements are presented. The system consists of three Dosepix detectors, which is a hybrid, pixelated, photon-counting X-ray detector. The energy and angular dependence of the normalized response and the coefficients of variation of the personal dose equivalents $H_p(10)$ and $H_p(0.07)$ are determined on an International Organization for Standardization (ISO) water slab phantom in continuous reference photon radiation fields according to ISO 4037-1 and ISO 4037-3. The energy response is presented for the narrow spectra N-15 to N-300 and the radiation qualities S-Cs and S-Co for angles of incidence of 0°, 30°, and ±60°. The highest deviation of the response from the reference response is found at N-60, -60° with 1.179 ± 0.007 (standard deviation) for $H_p(10)$. The energy range of use for $H_p(10)$ is expected to extend from 12.4 to 1250 keV. For the personal dose equivalent $H_p(0.07)$, the normalized response at ±60° is below the lower limit for N-15, N-20, and N-25. It results in an energy range of use from 24.6 to 1250 keV. The coefficient of variation increases with increasing photon energy and stays below 1% for all measurements.

Introduction

Active personal dosemeters (APDs) are critical for real-time radiation monitoring in environments such as interventional radiology and nuclear facilities. Traditional passive dosemeters lack immediate feedback, while many current APDs struggle with pulsed radiation fields and maintaining consistent accuracy across a broad energy and angular spectrum. This study evaluates a new dosimetry system based on the Dosepix hybrid pixel detector, which utilizes spectroscopic dose reconstruction to overcome these limitations.

Methodology

The experimental system comprises three Dosepix detectors mounted on a single readout board, each featuring a 300-μm-thick silicon sensor. To ensure accurate measurements across a massive energy range (12.4 keV to 1250 keV), the system employs a triple-filter design. As shown in Figure 1, each detector is covered by a different filter cap: an aluminum cylinder for soft X-rays, an aluminum half-sphere for medium energies, and a tin half-sphere for hard X-rays and gamma rays.

Schematic of the Dosepix detector system with triple-filter caps (Al and Sn) and PMMA housing.

The Dosepix ASIC operates in an energy binning mode, sorting photon events into 16 programmable energy bins. The total dose is calculated by weighting the counts in each bin with specific conversion factors. Testing was conducted at the Physikalisch-Technische Bundesanstalt (PTB) using an ISO water slab phantom and standard reference radiation qualities (N-series, $^{137}\text{Cs}$, and $^{60}\text{Co}$) at angles of $0^\circ$, $30^\circ$, and $\pm60^\circ$.

Results and Discussion

The spectroscopic reconstruction method achieved a remarkably flat energy response for the deep dose $H_p(10)$. At a $0^\circ$ angle of incidence, the mean normalized response was $0.998 \pm 0.011$, indicating high accuracy.

Normalized response for the personal dose equivalent $H_p(10)$ across a wide energy range and various angles of incidence.

As illustrated in Figure 2, the system maintains a stable response across the tested energy range. However, a slight overestimation (up to 1.179) was observed at steep angles ($\pm60^\circ$) for $H_p(10)$. This is attributed to the physical decrease in the air kerma-to-dose conversion factor at these angles while the detector continues to see a high photon fluence. For the shallow dose $H_p(0.07)$, the effective energy range was determined to be 24.6 keV to 1250 keV, as the response fell below regulatory limits at very low energies and steep angles. Precision remained excellent, with the coefficient of variation staying below 1% for all measurements, well within regulatory standards.

Conclusion

The Dosepix-based system demonstrates robust performance as an active personal dosemeter, meeting PTB requirements for a wide range of photon energies. Future developments will focus on optimizing filter geometries to improve the low-energy angular response for $H_p(0.07)$ and characterizing the system within a final wearable housing for legal certification.

Summary updated

Abstract

First measurements characterizing dosimetric properties of a dosimetry system designed for the purpose of active personal dosimetry for photons with mean energies from 12.4 to 1250 keV according to Physikalisch-Technische Bundesanstalt (PTB) requirements are presented. The system consists of three Dosepix detectors, which is a hybrid, pixelated, photon-counting X-ray detector. The energy and angular dependence of the normalized response and the coefficients of variation of the personal dose equivalents H <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">p</sub> (10) and H <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">p</sub> (0.07) are determined on an International Organization for Standardization (ISO) water slab phantom in continuous reference photon radiation fields according to ISO 4037-1 and ISO 4037-3. The energy response is presented for the narrow spectra N-15 to N-300 and the radiation qualities S-Cs and S-Co for angles of incidence of 0°, 30°, and ±60°. The highest deviation of the response from the reference response is found at N-60, -60° with 1.179 ± 0.007 (standard deviation) for H <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">p</sub> (10). The energy range of use for H <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">p</sub> (10) is expected to extend from 12.4 to 1250 keV. For the personal dose equivalent H <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">p</sub> (0.07), the normalized response at ±60° is below the lower limit for N-15, N-20, and N-25. It results in an energy range of use from 24.6 to 1250 keV. The coefficient of variation increases with increasing photon energy and stays below 1% for all measurements.
BibTeX
@article{haag2021personal,
  author = {Dennis Haag and Sebastian Schmidt and P. Hufschmidt and Franziska Eberle and Thilo Michel and G. Anton and Oliver Hupe and Jürgen Röth and Christian Fuhg and Hayo Zutz and R. Behrens and M. Campbell and X. Llopart and R. Ballabriga and L. Tlustos and W. C. Wong},
  title = {{Personal Dosimetry in Continuous Photon Radiation Fields With the Dosepix Detector}},
  journal = {IEEE Transactions on Nuclear Science},
  year = {2021},
  doi = {10.1109/tns.2021.3068832},
}
Figure