Abstract
The PS-DRF system offers a convenient, effective, and adaptable approach by integrating navigation technology with 3D printing. This method has the potential to simplify navigation-assisted mandibular surgery and facilitate the broader clinical implementation of computer-assisted navigation in maxillofacial procedures.
Introduction
Computer-assisted navigation (CAN) serves as a real-time "GPS" for surgeons, yet its application in mandibular (lower jaw) surgery is complicated by the bone's independent movement via the temporomandibular joint. Traditional tracking methods are either invasive, requiring screws to be fixed directly into the bone, or inaccurate, relying on shifting skin-surface markers. These challenges often lead to increased surgical trauma, longer procedure times, and a steep learning curve for clinicians.
To address these issues, researchers have developed a novel, non-invasive Patient-Specific Dynamic Reference Frame (PS-DRF). This system utilizes 3D printing to create a custom-fit device that attaches directly to the patient's teeth, allowing for precise, real-time tracking of the mandible without invasive fixation.
Methods
The PS-DRF workflow integrates advanced digital modeling and additive manufacturing:
- Digital Modeling and Design: A physical dental cast of the patient’s jaw is laser-scanned to create a high-fidelity 3D model. Using specialized software, a thin occlusal splint is designed to "snap" onto the teeth. This splint includes a rigid connective rod and a unique structure to hold a metal frame with infrared-reflective spheres.
- 3D Printing: The splint is fabricated using medical-grade resin on an Objet260 Connex3 3D printer, ensuring high precision (up to 200 microns).
- Registration: The design incorporates 1mm deep conical indentations as fiducial markers. During surgery, these markers provide a "lock-and-key" fit for the navigation probe, enabling rapid and accurate alignment between the virtual surgical plan and the physical patient.
Results and Clinical Validation
The technical method was validated through a clinical case involving a 29-year-old female patient with a right-sided condylar osteochondroma. Preoperative 64-slice CT scans were used for virtual surgical planning (VSP).
Key findings include:
- High Accuracy: The surgical outcome showed a deviation of less than 2 mm from the virtual simulation, meeting the strict requirements for maxillofacial procedures.
- Enhanced Efficiency: The PS-DRF reduced the registration and setup time to approximately 5 minutes, compared to the 10–20 minutes typically required for bone-fixed frames.
- Dynamic Tracking: Because the reference frame is fixed to the teeth, the navigation system remains synchronized with the jaw's movement, eliminating the need for re-calibration if the patient's position shifts.
Discussion and Limitations
The PS-DRF system significantly simplifies the workflow for navigation-assisted mandibular surgery. By replacing invasive bone screws with a "snap-on" splint and using pre-designed conical fiducials, the method reduces human error and surgical trauma.
However, certain limitations remain. The system requires stable dentition and sufficient mouth opening for dental impressions, making it unsuitable for edentulous patients or those with severe trismus. While this single-case study demonstrates clinical success, larger cohort studies are necessary to statistically validate the system's reliability across various pathologies. Future developments may include AI-driven anatomical recognition to further automate the planning phase.
Conclusion
The PS-DRF represents a significant step toward making high-precision, computer-assisted mandibular surgery more accessible and less invasive. By combining 3D printing with innovative design, this method offers a faster, more patient-friendly alternative to traditional navigation techniques.
Summary updated