OC18.01: Fusion of multiple four dimensional fetal echocardiography images can improve quality

Mark J. Gooding, Kashif Rajpoot, Salli Mitchell, P. Chamberlain, J. Alison Noble, Stephen KennedyUltrasound 2009 · DOI

Abstract

Four dimensional (4D) fetal echocardiography images, acquired using Spatio-temporal image correlation (STIC), can include imaging artefacts such as acoustic shadowing or reverberation. 4D fusion echocardiography (4DFE), a process whereby multiple images are combined, has been shown to improve image quality for adults. The aim of this study was to assess whether 4DFE is beneficial for fetal scanning, and which fusion method best improves image contrast while reducing image artefacts and speckle. Multiple 4D images were acquired for each fetus, with small orientation variations. The images in each set were rigidly aligned using an automatic tool. Five methods for 4DFE were compared, three averaging based methods: mean, median and mode, and two contrast enhancing methods: maximum and wavelet. The image contrast, signal-to-noise-ratio (SNR) and contrast-to-noise-ratio (CNR) were measured to assess the improvement in fused image quality in comparison to the individual 4D images. Multiple STIC scans for 7 volunteers were evaluated, with 3–7 (mean 3.7) 4D images used for fusion. Gestational ages ranged from 23 to 30 weeks. Fusion by averaging reduced speckle and imaging artefacts, as shown by an increase in SNR in the table. The contrast enhancing methods also enhanced speckle and reverberation artefacts, reducing SNR. All methods improved CNR, either by improving contrast or by reducing speckle. Qualitatively median intensity fusion seemed to offer the best improvement in image quality. Fusion of multiple images improved image quality and reduced the appearance of image artefacts. 4DFE using the median intensity appears to offer the best compromise between enhancing contrast and reducing speckle. OC18.01: Table
BibTeX
@inproceedings{gooding2009oc,
  author = {Mark J. Gooding and Kashif Rajpoot and Salli Mitchell and P. Chamberlain and J. Alison Noble and Stephen Kennedy},
  title = {{OC18.01: Fusion of multiple four dimensional fetal echocardiography images can improve quality}},
  booktitle = {Ultrasound},
  year = {2009},
  doi = {10.1002/uog.6562},
}

Loading related content...

Figure