Abstract
Inferior vena cava filter (IVCF) implantation is a standard procedure to prevent acute pulmonary embolism (PE) by capturing migrating thrombi. However, conventional filters often face issues with thrombus accumulation and mechanical instability. This study proposes a novel generation of bio-inspired helical IVCFs designed to induce helical blood flow. Using ANSYS 2023 for fluid-structure interaction (FSI) simulations and in vitro bench testing, we evaluated four helical geometries. Results indicate that helical structures significantly increase wall shear stress (WSS), promoting clot dissolution while enhancing mechanical stability against the vessel wall. The Expansion Pattern Filter (EPF) emerged as the most balanced design, offering effective trapping with minimal risk of migration or vessel perforation.
Introduction
Pulmonary embolism (PE), typically resulting from deep vein thrombosis (DVT) in the lower limbs, remains a major cause of cardiovascular mortality. IVCFs act as a mechanical barrier to "catch" these clots. Traditional designs, however, present a "catch-22": dense filters capture small clots but risk total blockage, while sparse filters may allow dangerous emboli to pass. Furthermore, mechanical forces can cause filters to tilt, migrate, or puncture the vena cava.
Inspired by the natural helical flow in the human arterial system, this research investigates whether spiral filter geometries can induce a "tornado effect" to improve both hemodynamics and mechanical anchoring.
Methodology
The study utilized a dual-track approach combining computational modeling and physical testing:
- Numerical Simulation: High-fidelity FSI simulations were conducted using ANSYS 2023. Blood was modeled as a non-Newtonian fluid using the Carreau model to capture realistic flow behavior. Four distinct geometries were tested: EPF (Expansion Pattern), CPF (Conical Pattern), SPF (Shrink Pattern), and LPF (Linear Pattern control).
- In Vitro Experiments: ABS plastic filters were 3D-printed and tested in a transparent resin IVC model. A peristaltic pump simulated physiological flow, and various microemboli (nylon beads and plush balls) were introduced to measure trapping efficiency and device displacement.
Hemodynamic and Mechanical Performance
The core innovation lies in the induction of helical flow dynamics. The simulations confirmed that helical structures successfully create swirling patterns, which increase Wall Shear Stress (WSS). This increased stress acts as a natural scouring mechanism, facilitating the "cleavage" or gradual dissolution of trapped clots.
Key Findings:
- Stability vs. Trapping: The SPF was most effective at capturing 2mm particles due to its dense central rods but experienced higher drag. Conversely, the EPF provided the best mechanical stability, exhibiting the lowest drag force ($0.93 \times 10^{-3}$ N) and causing the least vessel deformation.
- Bio-Inspired Safety: The curved contact surfaces of the helical designs distribute stress more uniformly. This reduces the risk of vessel wall perforation compared to the sharp struts of traditional filters.
- Control Failure: In physical tests, the LPF control group failed entirely, dislodging under flow, which underscores the superior "grip" provided by helical architectures.
Conclusion
The study concludes that the Expansion Pattern Filter (EPF) offers the most optimal balance between hemodynamic performance and mechanical safety. While the SPF excels in capture efficiency, the EPF’s superior stability and reduced risk of vessel injury make it a more viable candidate for clinical application. Future work will focus on transitioning to pulsatile flow models and conducting in vivo studies to evaluate biological responses at the helical contact points.
Reference: Huang, Y.X., Li, Q., Liu, M. et al. Numerical simulation and in vitro experimental study of the hemodynamic performance of vena cava filters with helical forms. Sci Rep 14, 17903 (2024).