Novel 1 H/ 19 F double‐tuned coil using an asymmetrical butterfly coil

Suk‐Min Hong, Chang‐Hoon Choi, Jörg Felder, N. Jon ShahMedical 2025 · DOI

TL;DR

This paper introduces a novel asymmetrical butterfly coil design for simultaneous 1H and 19F MRI. This new coil achieves higher signal-to-noise ratio (SNR) and better efficiency compared to conventional designs, overcoming key challenges in dual-nucleus MRI.

What Problem Does This Paper Address?

Magnetic Resonance Imaging (MRI) is a powerful tool for visualizing the human body. While proton (1H) MRI provides excellent anatomical detail, imaging other nuclei, like fluorine-19 (19F), offers complementary information. 19F MRI is particularly promising for applications like cell tracking and assessing tissue oxygenation, as it's non-invasive and doesn't produce background signals from the body. However, 19F MRI faces a significant hurdle: its signal-to-noise ratio (SNR) is much weaker than 1H MRI. This necessitates combining 19F coils with 1H coils for anatomical co-registration and shimming. A major challenge in creating these dual-tuned coils is the strong electromagnetic coupling between the coils, exacerbated by the close Larmor frequencies of 1H and 19F. Existing double-tuned coil methods often suffer from signal loss or require complex circuitry.

Methods

The researchers developed an "asymmetrical butterfly coil" designed to operate at both 1H and 19F frequencies. The key innovation lies in tuning the left and right loops of the butterfly coil asymmetrically. This asymmetry creates a "split resonance peak," allowing the coil to efficiently cover both the 1H and 19F frequencies without the need for lossy components like traps or switching circuits.

The coil's performance was evaluated using two main approaches:

  • Simulations: Finite Integration Technique (FIT) simulations were performed using CST Microwave Studio to model the coil's tuning and receive sensitivity. This involved loading the simulated coils with a cylindrical phantom.
  • MR Measurements: The constructed coils (both the asymmetrical butterfly and a conventional single-tuned loop coil for comparison) were tested on a 7T MRI scanner. 2D gradient echo images were acquired at both 1H and 19F frequencies. SNR maps were generated and analyzed.

The researchers compared the asymmetrical butterfly coil's performance against a dimension-matched single-tuned loop coil, focusing on transmission efficiency, receive sensitivity, and quadrature efficiency.

Datasets

Not discussed in the paper. The study utilized simulated phantoms and MR measurements on a 7T MRI scanner.

Key Insights and Ideas

  • Asymmetrical Tuning is Key: The core insight is that by deliberately tuning the butterfly coil's loops asymmetrically, a single coil can efficiently resonate at two distinct frequencies (1H and 19F) without significant signal loss. This is a departure from traditional symmetrical designs.
  • Split Resonance for Dual-Tuning: The asymmetrical tuning leads to a "split resonance peak," which is crucial for covering the separate frequencies of 1H and 19F. This avoids the need for complex switching mechanisms.
  • Improved SNR and Efficiency: The asymmetrical butterfly coil demonstrated superior receive sensitivity and SNR compared to a conventional single-tuned loop coil, making it more effective for 19F imaging.
  • No Additional Lossy Components: Unlike some previous dual-tuned coil designs that rely on PIN diodes or trap circuits, this asymmetrical butterfly coil achieves dual-tuning without these extra components, thus minimizing signal loss.

Key Results

  • Reflection Coefficients: Simulated results showed the asymmetrical butterfly coil achieved reflection coefficients (S11) better than -20 dB at both 1H and 19F frequencies, indicating good impedance matching. Conventional single-tuned coils struggled to cover both frequencies simultaneously.
  • Transmission Efficiency and Receive Sensitivity: Simulations indicated that the asymmetrical butterfly coil provided higher averaged receive sensitivity than the single-tuned loop coil at both 1H and 19F frequencies.
  • Quadrature Efficiency: The asymmetrical butterfly coil exhibited higher quadrature efficiency compared to the conventional loop coil.
  • Measured SNR: MR measurements confirmed that the asymmetrical butterfly coil provided a significantly higher averaged SNR than the single-tuned loop coil at both 1H and 19F frequencies. For example, at 19F frequency, the mean SNR was 23.4 for the asymmetrical butterfly coil versus 17.2 for the loop coil. At 1H frequency, the asymmetrical butterfly coil achieved a mean SNR of 532.2, compared to 378.2 for the loop coil.

Limitations and Future Work

  • Transmit Capability: While the coil is designed for dual-tuning, the paper notes that using it for transmission at both 1H and 19F frequencies might be limited due to local Specific Absorption Rate (SAR) concerns. Further evaluation of its transmit capabilities and potential modifications for optimal performance in transmission is suggested.
  • Phased Array Design: The current study focuses on a single coil. For whole-brain imaging or larger structures, a multi-channel phased array design using asymmetrical butterfly coils would be necessary. The paper suggests that existing decoupling methods for phased arrays could be adapted, but further research is needed to evaluate their effectiveness with this new coil design, particularly regarding bandwidth limitations with preamp decoupling.
  • 50 Ω Termination: The paper mentions that using 50 Ω termination for the asymmetrical butterfly coil array should be evaluated in future studies.

Summary updated

Abstract

The split resonance peak of the asymmetrical butterfly coil supported both <sup>1</sup>H and <sup>19</sup>F frequencies, delivering a higher SNR than that of the single-tuned loop coil. Since the asymmetrical butterfly coil can cover ¹H and ¹⁹F frequencies without loss and provides higher efficiency than the reference single-tuned coil, it can be effectively utilized for ¹H/¹⁹F MRI applications.