Novel approach to long‐term monitoring of accelerator‐based boron neutron capture therapy
Abstract
Neutron doses can be accurately administered by complementing proton beam current measurements with the online neutron beam monitor. The online neutron beam monitoring technique allows monitoring fluctuations in the neutron beam intensity and tracking the degradation of the lithium target through the neutron target lifespan. Using a calibrated online neutron beam monitor, a prescribed dose can be administered in a manner similar to that in x-ray therapy, and the duration of neutron beam irradiation on the patient can be controlled in real time.
BibTeX
@inproceedings{takada2025novel,
author = {Masashi Takada and Natsumi Yagi and Kenzi Shimada and Ryo Fujii and Masaru Nakamura and Satoshi Nakamura and Susumu Katō and Tomoya Nunomiya and Kei Aoyama and Masakuni Narita and T. Nakamura},
title = {{Novel approach to long‐term monitoring of accelerator‐based boron neutron capture therapy}},
booktitle = {Medical},
year = {2025},
doi = {10.1002/mp.17699},
}Ask Althea to put this paper to the test.