Nonunique Mappings of Water and k-edge Contrast Agent Pathlengths in Single-Energy CT Material Decomposition

Joel Phillips, Bo Chen, Emil Y. Sidky, Zheng Zhang, Dan Xia, Xiong Pan2025 · DOI

Abstract

The goal of this work is to investigate the existence of nonunique mappings of fluence transmission to material pathlength in conventional computed tomography (CT) projection material decomposition. Previous studies have demonstrated that a unique inversion of the nonlinear dual-energy CT (DECT) imaging model is only guranteed if the Jacobian determinant of the system is nonvanishing. Projection based basis material decomposition, which traditionally requires measurements with two different x-ray spectra, can be done with a conventional CT dataset supplemented by a pathlength conservation constraint. The Jacobian determinant for the single-energy CT system is derived and data is simulated for an object containing water and either iodine (25 mg/ml) or gadolinium (39.5 mg/ml) contrast agent. Vanishing determinant values are identified for multiple source spectra which are realistic to clinical diagnostic x-ray tube settings. The result indicates nonunique mappings from tranmission fluence measurements to material pathlengths are likely to exist and can diminish the accuracy of reconstructed basis images.
BibTeX
@misc{phillips2025nonunique,
  author = {Joel Phillips and Bo Chen and Emil Y. Sidky and Zheng Zhang and Dan Xia and Xiong Pan},
  title = {{Nonunique Mappings of Water and k-edge Contrast Agent Pathlengths in Single-Energy CT Material Decomposition}},
  year = {2025},
  doi = {10.1109/nss/mic/rtsd57106.2025.11287762},
}