Non-invasive photoacoustic computed tomography of rat heart anatomy and function

Li Lin, Xin Tong, Susana Cavallero, Yide Zhang, Shuai Na, Rui Cao, Tzung K. Hsiai, Lihong V. WangLight 2023 · DOI

TL;DR

This paper presents a novel non-invasive 3D photoacoustic computed tomography (PACT) system capable of imaging the entire rat heart with high anatomical and functional detail in just 10 seconds. This advancement overcomes previous limitations in PACT for cardiac imaging, enabling the study of cardiovascular diseases in animal models and holding promise for clinical translation to human neonates.

What Problem Does This Paper Address?

Cardiovascular diseases are a leading cause of death, necessitating robust preclinical research to understand disease mechanisms. While various non-invasive imaging modalities exist for small animal models (e.g., echocardiography, MRI, CT), each has limitations in terms of resolution, speed, or contrast. Photoacoustic computed tomography (PACT) offers functional optical contrast with high speed and resolution, but has struggled with non-invasive cardiac imaging due to challenges like:

  • Optical Attenuation and Acoustic Distortion: The chest wall, ribs, and lungs scatter and absorb light and sound, obscuring deep cardiac structures.
  • High Hemoglobin/Myoglobin Absorption: These endogenous chromophores in the heart provide contrast but also limit light penetration.
  • Cardiac Motion: The beating heart requires real-time imaging or sophisticated motion-correction techniques to capture dynamic structures.

Previous PACT prototypes have not been able to visualize the entire cardiac anatomy and its dynamics non-invasively without surgical intervention.

Methods

The researchers developed a modified three-dimensional PACT (3D-PACT) platform to address these challenges. Key methodological innovations include:

  • Regulated Illumination and Detection:
    • Hemispherical Detection Aperture: A large, hemispherical ultrasonic transducer array with a wide view angle was used to capture photoacoustic signals from the heart from numerous directions, mitigating acoustic blockage from ribs and lungs.
  • 1064-nm Laser Wavelength: This wavelength was chosen for its reduced scattering in biological tissues compared to commonly used near-infrared wavelengths, allowing for deeper light penetration into the heart.
  • Uniform Illumination: The laser beam was expanded to cover the imaging object, ensuring consistent light delivery.
  • ECG-Gated Synchronization: The PACT system was synchronized with the electrocardiogram (ECG) of the rat. This allowed for time-gating, where data acquisition was triggered by specific phases of the cardiac cycle (e.g., R-wave for ventricular contraction). This effectively reduces motion artifacts and enables the reconstruction of cardiac images at different points within a single heartbeat.
  • Rapid Scanning: The system was designed for rapid scanning, acquiring data for the entire heart within 10 seconds.
  • Image Reconstruction and Processing: A dual-speed-of-sound back-projection algorithm was used for reconstruction. Post-processing included depth compensation, high-pass filtering, sparse 4D transform-domain collaborative filtering for denoising, and a Hessian-based Frangi vesselness filter to enhance blood vessel visualization.

Datasets

Not discussed in the paper. The study utilized live Sprague Dawley (SD) rats, Zucker obese and lean rats, and spontaneously hypertensive rats for imaging.

Key Insights and Ideas

  • Overcoming the "Chest Wall Barrier": The combination of a large hemispherical detection array and the 1064-nm wavelength proved effective in penetrating the chest wall and capturing detailed cardiac structures, a significant hurdle for previous PACT systems.
  • Dynamic Cardiac Imaging in 10 Seconds: The ECG-gating and rapid scanning capabilities allow for the visualization of the beating heart's anatomy and function within a clinically relevant timeframe, moving beyond static imaging.
  • Unveiling Subtle Pathological Differences: The high contrast and resolution of 3D-PACT allowed for the clear differentiation of cardiac anatomy and hemodynamics between healthy, hypertensive, and obese rat models, revealing changes in chamber size, wall thickness, and blood flow patterns.
  • Potential for Clinical Translation: The non-ionizing nature of PACT, coupled with its speed and resolution, makes it a promising candidate for future clinical applications, particularly for imaging the hearts of human neonates where radiation exposure is a concern.

Key Results

  • Non-invasive Whole-Heart Imaging: The 3D-PACT platform successfully visualized the entire rat heart anatomy and intracardiac hemodynamics non-invasively.
  • High Imaging Speed: Cardiac imaging was achieved within 10 seconds.
  • Anatomical and Functional Characterization: The system enabled the measurement of cardiac chamber size, myocardial wall thickness, and intracardiac blood flow dynamics.
  • Disease Model Differentiation: Distinct structural and functional changes were observed in hypertensive and obese rat models compared to healthy controls, including increased wall thickness and altered chamber volumes.
  • Vascular Hemodynamics: PACT revealed variations in blood flow within major vessels like the aorta, pulmonary artery, and coronary arteries, showing differences across disease states.

Limitations and Future Work

  • Snapshot vs. Real-time: While the system captures dynamics over a 10-second scan, it does not provide true real-time imaging (i.e., instantaneous snapshots).
  • Resolution Enhancement: The authors suggest using ultrasonic arrays with higher center frequencies (5-10 MHz) for even finer anatomical detail, acknowledging that this may increase acoustic attenuation and rib disturbance.
  • Increased Transducer Density: Implementing more arc arrays with more elements could improve azimuthal sampling and reduce scanning steps for faster imaging.
  • Data Acquisition and Processing: Modifications to data amplification and acquisition circuits could integrate ultrasonography for coregistered imaging.
  • Future Directions: The authors envision PACT as a valuable tool for preclinical research and clinical translation, particularly for cardiac imaging of human neonates, due to its non-ionizing nature, speed, and functional contrast.

Summary updated

Abstract

Abstract Complementary to mainstream cardiac imaging modalities for preclinical research, photoacoustic computed tomography (PACT) can provide functional optical contrast with high imaging speed and resolution. However, PACT has not been demonstrated to reveal the dynamics of whole cardiac anatomy or vascular system without surgical procedure (thoracotomy) for tissue penetration. Here, we achieved non-invasive imaging of rat hearts using the recently developed three-dimensional PACT (3D-PACT) platform, demonstrating the regulated illumination and detection schemes to reduce the effects of optical attenuation and acoustic distortion through the chest wall; thereby, enabling unimpeded visualization of the cardiac anatomy and intracardiac hemodynamics following rapidly scanning the heart within 10 s. We further applied 3D-PACT to reveal distinct cardiac structural and functional changes among the healthy, hypertensive, and obese rats, with optical contrast to uncover differences in cardiac chamber size, wall thickness, and hemodynamics. Accordingly, 3D-PACT provides high imaging speed and nonionizing penetration to capture the whole heart for diagnosing the animal models, holding promises for clinical translation to cardiac imaging of human neonates.
BibTeX
@inproceedings{lin2023non,
  author = {Li Lin and Xin Tong and Susana Cavallero and Yide Zhang and Shuai Na and Rui Cao and Tzung K. Hsiai and Lihong V. Wang},
  title = {{Non-invasive photoacoustic computed tomography of rat heart anatomy and function}},
  booktitle = {Light},
  year = {2023},
  doi = {10.1038/s41377-022-01053-7},
}
Figure