Non-invasive arterial blood pressure measurement and SpO2 estimation using PPG signal: a deep learning framework

Yan Chu, Kaichen Tang, Yu-Chun Hsu, Tongtong Huang, Dulin Wang, Wentao Li, Sean I. Savitz, Xiaoqian Jiang, Shayan ShamsBmc Medical Informatics Decis. mak 2023

TL;DR

This paper introduces a novel deep learning framework using Transformer models to non-invasively estimate arterial blood pressure (ABP) and oxygen saturation (SpO2) from photoplethysmography (PPG) signals. The approach achieves high accuracy, meeting clinical standards, and offers a more convenient and less risky alternative to traditional methods, potentially improving continuous patient monitoring.

What Problem Does This Paper Address?

Monitoring blood pressure (BP) and oxygen saturation (SpO2) is critical for managing chronic diseases like hypertension and vascular disease. Current arterial blood pressure (ABP) measurement methods, while considered the gold standard, are invasive (e.g., arterial catheterization), carrying risks of discomfort, infection, and complications. Non-invasive methods like cuff-based devices can cause discomfort and introduce inaccuracies due to factors like patient movement. Photoplethysmography (PPG) signals, easily collected non-invasively, have shown promise for estimating BP. However, existing PPG-based BP estimation methods often rely on complex feature engineering, may struggle with varying signal lengths, and have been limited by small sample sizes and a lack of robustness. Furthermore, no previous study has simultaneously estimated both ABP and SpO2 using a single, end-to-end deep learning model from raw PPG signals. This paper aims to bridge this gap by developing a robust, non-invasive, and personalized deep learning framework for simultaneous ABP and SpO2 estimation from PPG.

Methods

The core of the proposed method is a Transformer-based deep learning architecture. Specifically, the authors employ a Time Series Transformer (TST) encoder. The raw PPG signals are first preprocessed, including filtering and Empirical Mode Decomposition (EMD) to extract features across multiple channels. These multi-channel time-series PPG signals are then fed into the Transformer encoder. The Transformer's self-attention mechanism is key, allowing it to capture temporal dependencies and internal patterns within the PPG signals without relying on fixed-size kernels or recurrent memory states, which can be computationally expensive and prone to information loss.

A significant aspect of their approach is the use of unsupervised pre-training followed by supervised fine-tuning. The unsupervised pre-training uses a Masked Language Modeling (MLM) strategy, where parts of the PPG sequence are masked, and the model learns to predict them. This helps the model learn general latent patterns from a diverse cohort. Subsequently, supervised fine-tuning is applied for personalized prediction, where the pre-trained model is adapted to individual patient data. This two-stage approach aims to improve robustness, generalizability, and personalized accuracy. The model is designed as an end-to-end system, directly processing raw PPG signals to output ABP (systolic and diastolic) and SpO2 values.

Datasets

The study utilizes the MIMIC III database, a publicly available critical care dataset. Specifically, they extracted waveform data from a subset of patients containing both ABP and PPG signals. After rigorous preprocessing and quality checks, a final dataset of 1,732 subjects with a total of 804,676 PPG slices was used for training, validation, and testing. This dataset is noted as being significantly larger than those used in many previous PPG-based BP estimation studies.

Key Insights and Ideas

  • End-to-End Transformer for PPG: The application of a Transformer encoder directly to raw PPG signals for simultaneous ABP and SpO2 estimation is a novel approach, moving beyond traditional feature engineering.
  • Unsupervised Pre-training for Robustness: Using MLM for unsupervised pre-training on a large, diverse dataset helps the model learn robust representations of PPG signals, making it less susceptible to noise and variations.
  • Personalized Fine-tuning for Accuracy: The fine-tuning strategy, adapted to individual patient data, significantly boosts prediction accuracy, demonstrating the importance of personalized models in healthcare.
  • Simultaneous Estimation Advantage: The ability to predict both ABP and SpO2 from a single PPG signal stream simplifies the monitoring process and reduces the need for multiple devices.
  • Scalability and Hardware Independence: The reliance on PPG signals, which can be acquired from readily available sensors, makes the approach hardware-independent and potentially scalable for widespread use.

Key Results

The proposed MLM-Transformer with personalization achieved impressive results on the MIMIC III test set:

  • Arterial Systolic Blood Pressure (ASBP): Mean Absolute Error (MAE) of 2.41 ± 2.72 mmHg.
  • Arterial Diastolic Blood Pressure (ADBP): MAE of 1.31 ± 1.77 mmHg.
  • Oxygen Saturation (SpO2): MAE of 0.56 ± 0.79%.

These results meet the stringent requirements of the Association for the Advancement of Medical Instrumentation (AAMI) standard and achieve "Grade A" according to the British Hypertension Society (BHS) grading standard. The model also demonstrated superior performance compared to baseline Transformer and U-Net models, as well as other state-of-the-art methods, particularly in terms of MAE.

Limitations and Future Work

The authors acknowledge several limitations:

  • Data Bias: PPG signals can be sensitive to skin color and other physiological variations, potentially leading to bias in the collected data and affecting generalizability to diverse populations.
  • Comorbidities: The study did not investigate the relationship between comorbidities and prediction performance, which could be a factor in clinical outcomes.
  • Generalizability to Non-ICU Cohorts: The MIMIC III dataset primarily consists of ICU patients. Further validation on external datasets, including those from healthy individuals, is needed to assess broader generalizability.
  • Computational Cost of Personalization: While personalization improves accuracy, it can be computationally expensive, posing a challenge for real-time deployment.
  • Limited Demographic Analysis: While some subgroup analysis was performed, a more in-depth investigation into demographic influences could be beneficial.

Future work suggested includes:

  • Evaluating on Diverse Datasets: Testing the model on datasets from healthy individuals and other patient cohorts.
  • Incorporating Additional Data: Integrating demographic features, clinical information, or other physiological signals to further enhance model performance.
  • Optimizing Personalization: Developing more computationally efficient methods for personalized model deployment.
  • Exploring Different Loss Functions: Investigating loss functions tailored for multi-task learning to better handle the heterogeneity of ABP and SpO2 data.
  • Real-world Deployment: Conducting studies in clinical settings to evaluate the model's practical utility and impact.

Summary updated

BibTeX
@article{chu2023non,
  author = {Yan Chu and Kaichen Tang and Yu-Chun Hsu and Tongtong Huang and Dulin Wang and Wentao Li and Sean I. Savitz and Xiaoqian Jiang and Shayan Shams},
  title = {{Non-invasive arterial blood pressure measurement and SpO2 estimation using PPG signal: a deep learning framework}},
  journal = {Bmc Medical Informatics Decis. mak},
  year = {2023},
  doi = {10.1186/s12911-023-02215-2},
}
Figure