NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

Non-contact physiological monitoring of post-operative patients in the intensive care unit

João Jorge, Mauricio Villarroel, Hamish Tomlinson, Oliver Gibson, Julie L. Darbyshire, Jody Ede, Mirae Harford, J. Duncan Young, Lionel Tarassenko, Peter J. WatkinsonNpj Digit. Medicine 2022

TL;DR

This paper introduces and validates a non-contact video camera system for continuous monitoring of heart rate (HR) and respiratory rate (RR) in post-operative intensive care unit (ICU) patients. The system demonstrates feasibility and accuracy comparable to traditional bedside monitors, offering a less intrusive alternative that could improve patient outcomes by enabling earlier detection of physiological deterioration.

What Problem Does This Paper Address?

Patients recovering in the ICU after major surgery are at a high risk of adverse events. Early detection of physiological deterioration is crucial for preventing these events and improving outcomes. Current monitoring relies on vital sign measurements taken manually at intervals or continuously via contact-based sensors. Manual measurements can miss subtle changes, while contact sensors are uncomfortable, limit patient mobility and rehabilitation, and are prone to issues like sensor disconnection. This study addresses the need for a reliable, non-intrusive method for continuous vital sign monitoring in the ICU that can capture early signs of deterioration without compromising patient comfort or mobility.

Methods

The core of this study is a non-contact monitoring system that uses a video camera to extract physiological signals. The system leverages Photoplethysmographic Imaging (PPGi), which analyzes changes in light reflected from the skin to derive vital signs.

Key methodological aspects include:

  • PPGi Signal Extraction: The system captures video data and processes it to extract PPGi signals from multiple regions of interest (ROIs) on the patient's skin.
  • Heart Rate (HR) Estimation: A pipeline, validated in previous studies, is used. It involves identifying salient points in the PPGi signal, assessing signal quality, and then using autoregressive spectral analysis and a Kalman filter for robust HR estimation. Estimates are reported every second.
  • Respiratory Rate (RR) Estimation: Similar to HR, respiratory signals are acquired by measuring intensity changes in ROIs related to chest movements. Data fusion algorithms and adaptive peak/onset detection are employed, with estimates reported every 5 seconds.
  • Gold-Standard Reference: To ensure accuracy, the system's estimates were compared against data from standard, clinically-approved bedside monitors (ECG for HR, impedance pneumography for RR, and pulse oximetry for HR). A rigorous process was used to identify and discard poor-quality reference data, ensuring the "valid time" used for comparison was robust.
  • Privacy Blind: A crucial component for ethical considerations in an ICU setting, a privacy blind was integrated into the camera system, allowing staff or patients to obscure the view when necessary.

Datasets

The study utilized data collected from 15 post-operative patients admitted to the ICU at the Churchill Hospital in Oxford, UK. Recordings were taken over clinically relevant periods, with an average session duration of approximately 15.6 hours, encompassing both day and night. This dataset represents a novel benchmark for non-contact vital sign monitoring in an acute ICU setting over extended durations, differing from previous studies that often focused on shorter periods or less critical environments.

Key Insights and Ideas

  • Feasibility in a Challenging Environment: The study successfully demonstrates that non-contact vital sign monitoring is feasible and robust even in the demanding ICU environment, which presents challenges like variable lighting and potential patient movement.
  • Night-time Performance is Comparable: Contrary to potential assumptions, the system performed comparably, and in some metrics even better, during night-time monitoring, suggesting that low-light conditions do not significantly hinder its effectiveness.
  • Early Detection Potential: The ability to capture subtle, transient changes in vital signs, as demonstrated in case studies of patient deterioration, highlights the system's potential for early warning of clinical decline.
  • Reduced Patient Burden: By eliminating the need for contact sensors, the system offers a significant advantage in patient comfort and mobility, which is crucial for rehabilitation and overall recovery.
  • Beyond Vital Signs: The exploration of "localised respiration mapping" suggests that video-based monitoring could offer richer diagnostic information, potentially assessing respiratory effort and asymmetries, going beyond simple vital sign estimation.

Key Results

  • Heart Rate: The system achieved an estimated HR for 53.2% (103.0 hours) of the total valid camera data with a Mean Absolute Error (MAE) of 2.5 beats/min compared to reference sensors. The correlation coefficient was high at r = 0.98.
  • Respiratory Rate: RR was estimated for 63.1% (119.8 hours) of the total valid camera data with an MAE of 2.4 breaths/min. The correlation coefficient was r = 0.82.
  • Coverage: A total of 233.5 hours of recording data were collected. Valid time for HR estimation was 193.7 hours (82.9%), and for RR estimation was 189.7 hours (81.3%).
  • Day vs. Night: While overall performance was similar, night-time monitoring showed slightly lower MAEs for both HR (2.5 bpm vs. 2.7 bpm) and RR (2.3 brpm vs. 2.6 brpm), with higher correlation coefficients.
  • Privacy Time: The use of the privacy blind accounted for a manageable portion of the recording time, ranging from 1.4% to 41.8% per participant, indicating that dignity concerns can be addressed without completely sacrificing monitoring.

Limitations and Future Work

  • Single-Center Study: The data was collected at a single ICU in Oxford, UK. Generalizability to other ICUs, healthcare systems, or patient populations may be limited.
  • Patient Mobility: While less intrusive, the system's effectiveness with fully ambulatory patients in lower-acuity settings would require further investigation, potentially with wider-angle lenses.
  • Skin Tone Diversity: The study cohort was predominantly of lighter skin tones (93.3% Fitzpatrick skin type I or II). The performance of HR estimation algorithms can be affected by darker skin tones, necessitating validation on a more diverse cohort.
  • Patient Position: The impact of patient position on accuracy and coverage was not assessed.
  • Future Work: The authors suggest further research into capturing haemodynamic events, exploring the diagnostic potential of visual cues beyond vital signs, and adapting the system for less tethered patient environments. They also highlight the need for validation across diverse patient demographics and clinical settings.

Summary updated

BibTeX
@article{jorge2022non,
  author = {João Jorge and Mauricio Villarroel and Hamish Tomlinson and Oliver Gibson and Julie L. Darbyshire and Jody Ede and Mirae Harford and J. Duncan Young and Lionel Tarassenko and Peter J. Watkinson},
  title = {{Non-contact physiological monitoring of post-operative patients in the intensive care unit}},
  journal = {Npj Digit. Medicine},
  year = {2022},
  doi = {10.1038/s41746-021-00543-z},
}
Figure