Abstract
The Jugular Venous Pulse (JVP) is a critical indicator of right atrial heart pressure, yet its measurement typically requires invasive catheterization or specialized contact-based ultrasound. This paper presents Coded Hemodynamic Imaging (CHI), a non-contact optical system capable of capturing the JVP waveform from a distance. By analyzing near-infrared light-tissue interactions, CHI distinguishes between arterial and venous flow, providing a "touchless" window into cardiac health.
Introduction
The Jugular Venous Pulse (JVP) serves as a direct proxy for the heart’s right atrium. Distortions in the JVP waveform can indicate serious conditions such as heart failure, tricuspid stenosis, or arrhythmias. Despite its clinical value, JVP monitoring is rarely routine because the gold standard—central venous catheterization—is invasive. Non-invasive alternatives like Doppler ultrasound require physical contact and expert operation. This research introduces a "touchless" method to monitor these vital hemodynamic signals, potentially streamlining cardiac assessments in clinical settings.
Methods: Coded Hemodynamic Imaging (CHI)
The proposed system utilizes Coded Hemodynamic Imaging (CHI), an advanced form of photoplethysmographic imaging (PPGI). Unlike standard PPGI, CHI is designed to resolve complex spatiotemporal blood flow patterns across a wide field of view.
Hardware and Setup
The system consists of a near-infrared (NIR) camera (850–1000 nm) and a 250W tungsten-halogen light source with a spatial diffuser for uniform illumination. Positioned approximately 1.5 meters from a supine participant, the camera captures reflected light which is then converted into absorbance changes using the Beer-Lambert law.
Signal Processing
The neck area is analyzed in 5x5 mm regions. To differentiate between vessel types, the system calculates the Pearson’s linear correlation between local neck signals and a reference finger pulse sensor. This allows the identification of arterial flow (synchronized with the finger pulse) and venous flow (often exhibiting an inverse relationship).
Key Insights and Results
The study, involving 24 participants, demonstrated that CHI can successfully detect both arterial and venous waveforms non-invasively.
The "Inverted" Venous Pulse
A primary finding was the bimodal distribution of signals in the neck. While the carotid artery pulses in sync with systole, the jugular vein exhibits an "inverted" pulse relative to the arterial signal, with a strong inverse correlation ($\rho = -0.96$).
Waveform Fidelity and Morphology
CHI captured the specific morphology of the JVP waveform, identifying the classic $a, c, x, v,$ and $y$ waves corresponding to mechanical cardiac events. The system also mapped the spatial trajectory of the jugular vein, showing how the pulse wave travels across the neck.
Anatomical Validation
The locations identified by CHI as venous were consistently confirmed by B-mode/Doppler ultrasound to be on the distal side of the carotid artery, matching known human anatomy.
Discussion and Future Work
While CHI successfully observed waveforms in all participants, signal quality can be affected by body fat content or skin inelasticity, which dampen surface movements. Additionally, respiratory-induced changes in intrathoracic pressure introduce variability that must be addressed in future automated tools.
Future research will focus on clinical validation with patients suffering from cardiac abnormalities, such as tricuspid regurgitation, to determine if CHI can accurately detect the specific waveform distortions used for diagnosis. This technology paves the way for "smart" clinical environments where cardiac health can be assessed passively by specialized camera systems.
Summary updated