Noise‐augmented deep denoising: A method to boost CT image denoising networks

Gernot Kristof, Elias Eulig, Marc KachelrießMedical 2025 · DOI

TL;DR

This paper introduces "Noise-Augmented Deep Denoising" (NADD), a novel method that significantly improves CT image denoising networks by providing them with simulated noise realizations as additional input. This approach helps networks better understand and remove noise artifacts, leading to higher quality images with reduced radiation dose.

What Problem Does This Paper Address?

Computed Tomography (CT) is a vital medical imaging tool, but its use of ionizing radiation raises concerns about patient safety. The "As Low As Reasonably Achievable" (ALARA) principle dictates minimizing radiation dose, which often leads to increased image noise, degrading image quality. While deep learning methods have shown promise in denoising CT images, conventional approaches struggle to effectively handle the complex, correlated noise inherent in CT scans. This noise is not random but arises from the global attenuation properties of the patient and the imaging system, often appearing as streaky artifacts. Existing deep denoising networks, with their limited receptive fields, often fail to capture this non-local noise structure, sometimes mistaking noise for anatomical features. This paper aims to bridge this gap by providing networks with explicit information about the noise characteristics.

Methods

The core innovation of this paper is the Noise-Augmented Deep Denoising (NADD) approach. Instead of solely relying on the low-dose CT (LDCT) image, NADD augments the input by generating N simulated noise-only realizations. These noise realizations are created by injecting Poisson noise into the raw projection data of a low-dose scan, reconstructing these noisy projections, and then subtracting the original (noise-free) reconstructed image to isolate the noise component.

These N noise-only images, along with the original LDCT image, are then fed as additional input channels into existing deep neural networks. The NADD method is architecture-agnostic, meaning it can be applied to various denoising network architectures. The authors demonstrate this by modifying three popular denoising networks: CNN10, ResNet, and WGAN-VGG. These modified networks are trained and used for inference with these additional noise channels. The rationale is that by explicitly providing information about the noise structure, the networks can learn to better distinguish between true anatomical structures and noise artifacts, leading to improved denoising performance.

Datasets

The study utilizes a dataset of 39 standard-dose clinical CT volumes acquired from a Siemens Somatom Force CT system, resulting in approximately 50,000 CT images. These volumes cover the neck to abdomen of patients. The dataset was split into training, validation, and test sets (27:8:4 ratio). For training and evaluation, low-dose CT (LDCT) images were generated by simulating a 90% dose reduction. The authors also generated "medium dose" images (25% dose reduction) to test the robustness of their trained networks.

Key Insights and Ideas

  • Noise is not random, it's structured: The paper emphasizes that CT noise is correlated and patient-specific, not simply random pixel-wise variations. This structured nature is key to understanding why NADD is effective.
  • Explicit noise information is powerful: Providing networks with simulated noise realizations acts as a strong prior, guiding the denoising process and enabling the network to learn the complex noise patterns.
  • NADD is a general enhancement: The method is architecture-agnostic, suggesting it can be a plug-and-play improvement for a wide range of existing CT denoising networks.
  • More noise realizations can be better: The study shows a trend where increasing the number of noise-only inputs (up to N=10) generally leads to improved denoising performance.
  • Robustness to noise levels: NADD-enhanced networks still perform better even when applied to images with slightly different noise levels than those used during training, indicating a degree of robustness.

Key Results

  • Significant performance improvement: In all tested cases, NADD significantly improved the denoising performance of the CNN10, ResNet, and WGAN networks.
  • Removal of misidentified artifacts: NADD successfully removed noise artifacts that were previously misinterpreted as anatomical structures by the original networks.
  • Quantitative gains: Table 1 shows substantial improvements in SSIM (Structural Similarity Index Measure), PSNR (Peak Signal-to-Noise Ratio), and VIF (Visual Information Fidelity) for all tested networks when using NADD. For example, CNN10+10 achieved an SSIM of 0.9560, PSNR of 43.1015, and VIF of 0.3059, compared to the baseline CNN10's SSIM of 0.9373, PSNR of 41.6372, and VIF of 0.2615.
  • Impact of N: For CNN10, increasing N from 0 to 10 consistently improved all metrics, with notable gains already visible at N=3.
  • Robustness demonstrated: Table 2 shows that NADD-enhanced networks still outperformed baseline networks when applied to medium-dose images, even though training was done on low-dose data.

Limitations and Future Work

  • 2D limitation: The current study is limited to 2D image denoising. The authors suggest that extending NADD to 3D denoising could lead to further improvements.
  • Simulated noise: The noise realizations are simulated, not measured. Future work could explore using measured noise for a more realistic simulation and investigate if retraining is necessary for different reconstruction parameters.
  • Inference time: NADD introduces a slight increase in inference time due to the additional reconstructions required for the noise-only inputs.
  • Human reader studies: The authors acknowledge that the ultimate evaluation of image quality should involve human reader studies, which were beyond the scope of this work.
  • Generalizability: Future investigations could explore whether NADD-trained networks are less prone to anatomical structure invariance and "hallucinations" compared to traditional counterparts.
  • Normal dose scans: While NADD can be applied to normal dose scans, the improvements are expected to be less dramatic as noise is less of a limiting factor in image quality.

Summary updated

Abstract

Providing additional simulated noise realizations helps to significantly improve the performance of CT image denoising networks.
BibTeX
@inproceedings{kristof2025noiseaugmented,
  author = {Gernot Kristof and Elias Eulig and Marc Kachelrieß},
  title = {{Noise‐augmented deep denoising: A method to boost CT image denoising networks}},
  booktitle = {Medical},
  year = {2025},
  doi = {10.1002/mp.18121},
}
Figure