New leak assembly based on fluidic nanochannels

Aiqing Zhu, Yongheng Zhao, Xudi Wang, Yu Wang, Wei Wei, Keqiang Qiu, Dong DongJournal of Vacuum Science & Technology A Vacuum Surfaces and Films 2016

Abstract

Fluidic nanochannels with a characteristic dimension of ∼280 nm were fabricated and designed as a leak assembly, where the nanochannels were formed on silicon wafers and enclosed with Pyrex® glass. The geometric dimensions were characterized by scanning electron microscopy, and the gas flow conductance of He and other heavy gases (N2, O2, and Ar) was measured, and its uncertainty estimated, by the difference method. The results indicated that the measured flow conductance values were 45% less than the calculated flow conductance values. For helium, molecular flow was shown to occur at pressures ranging from vacuum to atmospheric pressure. As a consequence of the well-defined geometry, the prediction of flow conductance could be achieved for various gas species.
BibTeX
@article{zhu2016new,
  author = {Aiqing Zhu and Yongheng Zhao and Xudi Wang and Yu Wang and Wei Wei and Keqiang Qiu and Dong Dong},
  title = {{New leak assembly based on fluidic nanochannels}},
  journal = {Journal of Vacuum Science \& Technology A Vacuum Surfaces and Films},
  year = {2016},
  doi = {10.1116/1.4960646},
}

Loading related content...