New generation of fixed points
Lists
Only you see this
Abstract
In this paper,the new techniques to implement new generation of fixed points are reported.Compared with the traditional techniques,the new techniques have some obvious advantages in their architectures,easy operating as well as high efficiency,and are especially suitable to verification for both the second class and industrial thermomenters.A new recurrence method of triple-point of water,without using cold-producing medium like liquid nitrogen and dry ice etc.,can form three-phase equilibrium of water automatically and makes it maintain a few ten hours.The recurrence set of the melting point of Ga,shelled by stainless steel,can provide a melting plateau that lasts a week and uncertainty of 0.0001℃.The frezzing points of pure metals for the second class calibration,including In,Sn,Zn and Al,have uncertainties between 0.001℃ and 0.005℃.
BibTeX
@inproceedings{chen2005new,
author = {Deming Chen},
title = {{New generation of fixed points}},
booktitle = {China},
year = {2005},