NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

Neuromorphic Heart Rate Monitors: Neural State Machines for Monotonic Change Detection

Alessio Carpegna, Chiara De Luca, Federico Emanuele Pozzi, Alessandro Savino, Stefano Di Carlo, Giacomo Indiveri, Elisa DonatiarXiv 2024 · DOI

TL;DR

This paper introduces a ultra-low-power neuromorphic Neural State Machine (NSM) designed to continuously monitor electrocardiogram (ECG) data and detect monotonic (unidirectional) increases in heart rate. Implemented on the mixed-signal DYNAP-SE neuromorphic processor, the system consumes a mere 90 $\mu$W of power. This makes it an ideal candidate for "wear and forget" health monitoring devices, particularly for detecting physiological stress and early signs of agitation in dementia patients.

What Problem Does This Paper Address?

Continuous, always-on health monitoring is vital for early medical intervention, but it poses severe engineering challenges. Wearable devices must process complex bio-signals in real time while operating under strict power constraints. This is especially critical for patients with cognitive impairments, such as dementia, who cannot be expected to regularly manage or charge their devices.

Tracking heart rate (HR) trends over extended periods can reveal early indicators of cardiac conditions (like arrhythmia or tachycardia) or psychological distress. However, traditional digital processing methods are often too power-hungry for continuous wear. The authors address this gap by leveraging neuromorphic engineering—hardware that mimics the brain's energy efficiency—to build a robust, real-time, and low-power heart rate monitor.

Methods

The researchers developed a Monotonic Neural State Machine (NSM) architecture and deployed it on the DYNAP-SE chip, a custom asynchronous mixed-signal neuromorphic processor featuring analog spiking neurons and synapses.

The pipeline consists of three main stages:

  1. Signal-to-Spike Conversion: Raw ECG signals are passed through four 4th-order Butterworth bandpass filters corresponding to different heart rate zones: 60–82 bpm (relaxed), 82–105 bpm, 105–128 bpm, and 128–150 bpm (tachycardiac). The filtered signals are full-wave rectified and converted into spike trains using Leaky Integrate-and-Fire (LIF) neurons.
  2. Winner-Take-All (WTA) Core: The spikes feed into four Excitatory-Inhibitory (EI)-balanced populations of Adaptive Exponential Integrate-and-Fire (AdEx I&F) neurons. A shared inhibitory population ensures a WTA dynamic, meaning only the state corresponding to the most active heart rate frequency band remains active.
  3. Monotonic Gating Mechanism: To ensure the network only tracks progressive increases in heart rate (ignoring temporary fluctuations or noise), the authors implemented a dis-inhibition mechanism. Gating populations continuously inhibit inactive higher states. When a specific state becomes active, it inhibits the gating population of the next state, effectively priming it to accept input.
Architecture of the Monotonic Neural State Machine

Datasets

The model was validated using a publicly available dataset containing physiological recordings from 8 participants. The data included chest ECG recordings taken during various physical activities, specifically indoor treadmill exercises (walking, jogging, running) and indoor cycling at varying resistance levels. The signals were sampled at 256 Hz and pre-processed with a 50 Hz notch filter to eliminate mains interference.

Key Insights and Ideas

  • Working Memory via EI Balance: A major challenge in neuromorphic design is processing slow biological signals using hardware neurons that have fast, millisecond-scale time constants. By organizing neurons into EI-balanced populations, the network achieves sustained activity (working memory) that outlasts individual neuron time constants, allowing it to track slow heart rate trends over minutes or hours.
  • Algorithmic Monotonicity via Dis-inhibition: Instead of relying on precise, fragile synaptic weight tuning, the network enforces unidirectional state transitions structurally through dis-inhibition. The system naturally ignores downward fluctuations or random noise spikes unless a full relaxation back to the baseline "State 0" occurs, which acts as a system reset.
Fig. 2: Network behavior when stimulated with 50Hz Poissonian sequences testing all the possible transitions: (a) non-monotonic WTA dynamics; (b) monotonic NSM response

Key Results

  • Hardware Efficiency: The entire network requires only 176 neurons (16 per population), easily fitting within a single core of the DYNAP-SE chip.
  • Ultra-Low Power: The system operates at an average firing rate of ~50 Hz per population, resulting in an incredibly low total power consumption of approximately 90 $\mu$W.
  • Real-World Validation: When tested against real ECG data, the network successfully distinguished between different activity profiles:
  • During an intense 10-minute cycling session, the network accurately tracked the progressive "ramp-up" of the heart rate, successfully triggering the final "Alarm" state while ignoring transient noise.
  • During a 6-minute walking session, the heart rate increase was only partial; the network correctly advanced to State 1 and safely held its state without falsely triggering the alarm.
Fig. 4: Response of the network when stimulated with a real ECG signal: (a) intense 10 minutes bike session; (b) 6 minutes and 40 seconds of walk

Limitations and Future Work

While the results are promising, the authors note a few limitations moving toward clinical deployment:

  • ECG vs. PPG: The current system was validated using chest-strap ECG data. For true wearable comfort (e.g., smartwatches), the system needs to be adapted to Photoplethysmogram (PPG) signals measured at the wrist.
  • Noise and Artifacts: Wrist-based PPG signals are highly susceptible to motion artifacts, which will require more sophisticated front-end filtering.
  • Clinical Patient Testing: The system has only been validated on healthy subjects performing exercise. Future studies must transition to testing patients with actual neuropathologies, such as dementia, to verify how the network handles real-world episodes of physiological agitation.

Summary updated

BibTeX
@misc{carpegna2024neuromorphic,