Abstract
NeuralOCT introduces a novel teacher-student framework utilizing neural fields to reconstruct high-fidelity 3D airway models from anatomic Optical Coherence Tomography (aOCT) scans. By bridging 2D deep-learning segmentation with implicit 3D representations, the method achieves sub-70 micrometer accuracy. This provides a radiation-free alternative to CT for assessing pediatric airway abnormalities like subglottic stenosis, offering a continuous, high-resolution geometry suitable for surgical planning and computational simulations.
Introduction
Imaging the tiny airways of infants and children is a high-stakes clinical challenge. Conditions such as subglottic stenosis—a life-threatening narrowing of the airway—require precise geometric measurements to determine whether a patient needs invasive surgery or can be managed through observation. While Computed Tomography (CT) scans provide detailed 3D data, they expose pediatric patients to harmful ionizing radiation.
Anatomic Optical Coherence Tomography (aOCT) has emerged as a promising radiation-free alternative. It utilizes a miniature fiber-optic catheter that traverses the airway, emitting a rotating laser to create a helical scan path. However, aOCT does not produce a standard 3D volume; instead, it generates a series of 2D "pull-back" frames. Converting these noisy, helical slices into a smooth, accurate 3D mesh is mathematically complex, and existing methods often struggle with noise or insufficient resolution for surgical planning.
Methodology: The Teacher-Student Framework
NeuralOCT addresses the reconstruction challenge by splitting the task into two distinct phases: a 2D segmentation "teacher" and a 3D neural field "student."
1. The Teacher (2D Segmentation)
The system first processes individual 2D OCT frames using nnU-Net, a state-of-the-art medical image segmentation framework. This "teacher" identifies the airway wall and extracts a raw 3D point cloud by converting the polar coordinates of the scan into a Cartesian (x, y, z) space.
2. The Student (3D Neural Field)
Rather than simply interpolating the point cloud, the "student" module employs a Neural Field. This coordinate-based neural network learns a Signed Distance Function (SDF), representing the airway surface as a continuous function where the surface is defined as the zero-level set.
3. NeuralPull Optimization
To train the student, the authors utilize the NeuralPull operator. This allows the network to "pull" arbitrary 3D query points toward the nearest point on the predicted surface. This process effectively denoises the raw 2D segmentations and creates a smooth, "infinite-resolution" 3D geometry that can be queried at any point in space.
Experimental Setup and Datasets
The study utilized a specialized dataset of 35 airway OCT scans, containing between 100 and 600 frames each. Expert clinicians manually segmented 931 frames to establish a ground truth gold standard. The data was split into 25 scans (819 frames) for training and 10 scans (112 frames) for testing. Images were preprocessed to a 1024x1024 resolution and normalized based on mean intensity and standard deviation.
Key Results
NeuralOCT demonstrated significant improvements over existing baselines:
- High Precision: The method achieved an average A-line error (distance error along the laser path) of less than 70 micrometers.
- Geometric Superiority: In 3D reconstruction benchmarks, NeuralOCT outperformed MedSAM and Local-NISF across all geometric metrics, including Chamfer Distance (CD) and Hausdorff Distance (HD).
- Denoising Capabilities: The neural field naturally filters out "jitter" and outliers produced by the 2D segmentation network by enforcing surface continuity.
- Efficiency: Training the implicit neural model for all test scans takes approximately one hour on a standard consumer GPU (NVIDIA 3090 Ti).
Interestingly, the specialized nnU-Net outperformed the foundation model MedSAM, suggesting that niche medical modalities still benefit significantly from domain-specific architectures over general-purpose "big data" models.
Conclusion and Future Work
NeuralOCT provides a robust pathway for high-resolution, radiation-free 3D airway imaging. By representing the airway as an implicit function rather than an explicit mesh, the model enables downstream applications like Computational Fluid Dynamics (CFD) simulations of airflow.
Future work will focus on jointly optimizing the teacher and student modules to allow 3D geometric priors to inform 2D segmentation in real-time. Additionally, expanding the dataset to include a wider variety of pediatric airway pathologies will be essential for clinical validation.
Summary updated