Navigation of guidewires and catheters in the body during intervention procedures: a review of computer-based models

Hoda Sharei, Tanja Alderliesten, John J. van den Dobbelsteen, Jenny DankelmanJournal of Medical Imaging 2018 · DOI

TL;DR

This paper reviews existing computer-based models for simulating the navigation of guidewires and catheters in the body. It categorizes these models by their purpose (training, planning, design) and the underlying techniques (FEM, MSM, rigid multibody links), highlighting their strengths and weaknesses for different applications. The review emphasizes the critical trade-off between simulation realism and computational efficiency, crucial for advancing these tools in medical practice.

What Problem Does This Paper Address?

Minimally invasive procedures, like those using guidewires and catheters, offer significant patient benefits but pose challenges for clinicians. Specialists lose direct tactile feedback and visual cues, relying on indirect methods like X-ray. Learning these complex manipulation skills traditionally involves risky, time-consuming patient interaction or expensive cadaver/animal models. Computer simulations offer a safer, more efficient alternative for training, pre-procedure planning, and even instrument design. However, a comprehensive understanding of the various modeling approaches, their suitability for different tasks, and their validation methods has been lacking, hindering their widespread adoption and improvement.

Methods

The authors conducted a literature review using Google Scholar and Web of Science, focusing on English-language publications from 1960 to 2017. They searched for keywords related to guidewire and catheter modeling, simulation, training, virtual reality, and vascular phantoms. The review systematically surveyed different modeling techniques, including:

  • Finite-Element Method (FEM): A robust but computationally intensive method that discretizes the instrument into small elements to model its deformation.
  • Mass-Spring Model (MSM): A simpler approach where the instrument is represented as a network of masses connected by springs and dampers, offering faster computation but potential instability.
  • Rigid Multibody Links: This method models the instrument as a series of rigid bodies connected by joints, offering a good balance of simplicity, speed, and ease of implementation, especially for real-time applications.
  • Hybrid Models: Combinations of the above techniques, often used to leverage the strengths of different methods for different parts of the instrument (e.g., a rigid body for the stiff shaft and MSM for a flexible tip).

The review also discusses factors influencing instrument-vessel interaction, such as collision and friction, and various validation strategies.

Datasets

Not discussed in the paper. The paper focuses on a review of existing literature and modeling techniques rather than presenting new experimental data or datasets.

Key Insights and Ideas

  • Purpose-Driven Modeling: The choice of modeling technique is heavily dictated by the intended application. Training simulations prioritize real-time performance, often at the expense of some realism, while design and planning might favor higher fidelity even with longer computation times.
  • The Realism vs. Speed Trade-off: This is the central challenge. Highly realistic models (e.g., FEM) are computationally expensive, making them unsuitable for real-time training. Simpler models (e.g., rigid multibody links) are faster but may sacrifice accuracy or the ability to capture complex phenomena.
  • Vessel Interaction is Key: Forces like contact with the vascular wall and friction are critical for realistic simulation. Accurately modeling these interactions is essential for predicting instrument behavior and providing meaningful haptic feedback.
  • Validation is Crucial but Inconsistent: While face validity (expert opinion) and comparison with phantom experiments are common validation methods, there's a lack of standardized experimental methodologies, making it difficult to compare the robustness of different models.
  • Assumptions are Necessary: Due to the inherent complexity of the vascular system and instrument mechanics, researchers often make simplifying assumptions about material properties, vessel geometry, and interaction forces.

Key Results

The paper provides a comprehensive overview and classification of existing modeling techniques for guidewires and catheters. It highlights the strengths and weaknesses of FEM, MSM, and rigid multibody links for different purposes:

  • FEM: High numerical stability, but very high computational effort.
  • MSM: Faster than FEM, but risks numerical instability.
  • Rigid Multibody Links: Simple structure, easy to implement, and potentially faster than MSM for longer instruments, but can become slow with many segments.

The review identifies that while some validated models are already in practical use, ongoing development is needed for further improvement, particularly in capturing complex vessel interactions and achieving consistent validation.

Limitations and Future Work

The authors acknowledge several limitations and suggest future research directions:

  • Vessel Deformation: Most studies assume rigid vessels, neglecting deformation, which can occur with stiff instruments or in certain pathologies. Future work should investigate vessel deformation more thoroughly.
  • Vessel Geometry: Current models often simplify vessels to circular cross-sections. Future research should explore more complex, patient-specific geometries and how they influence navigation.
  • Friction Modeling: While friction is important, its properties are often determined empirically, and its accurate modeling remains a challenge.
  • Validation Standardization: The lack of consistent experimental methodologies for validation is a significant limitation. Developing standardized protocols is crucial for comparing and advancing models.
  • Material Properties: Manufacturers rarely provide detailed mechanical properties for instruments, forcing empirical determination. More standardized characterization would improve model accuracy.
  • Real-time Performance: Achieving both high realism and real-time performance remains a significant challenge, especially for complex simulations. Hybrid approaches and algorithmic optimizations will likely be key.

Summary updated

Abstract

Guidewires and catheters are used during minimally invasive interventional procedures to traverse in vascular system and access the desired position. Computer models are increasingly being used to predict the behavior of these instruments. This information can be used to choose the right instrument for each case and increase the success rate of the procedure. Moreover, a designer can test the performance of instruments before the manufacturing phase. A precise model of the instrument is also useful for a training simulator. Therefore, to identify the strengths and weaknesses of different approaches used to model guidewires and catheters, a literature review of the existing techniques has been performed. The literature search was carried out in Google Scholar and Web of Science and limited to English for the period 1960 to 2017. For a computer model to be used in practice, it should be sufficiently realistic and, for some applications, real time. Therefore, we compared different modeling techniques with regard to these requirements, and the purposes of these models are reviewed. Important factors that influence the interaction between the instruments and the vascular wall are discussed. Finally, different ways used to evaluate and validate the models are described. We classified the developed models based on their formulation into finite-element method (FEM), mass-spring model (MSM), and rigid multibody links. Despite its numerical stability, FEM requires a very high computational effort. On the other hand, MSM is faster but there is a risk of numerical instability. The rigid multibody links method has a simple structure and is easy to implement. However, as the length of the instrument is increased, the model becomes slower. For the level of realism of the simulation, friction and collision were incorporated as the most influential forces applied to the instrument during the propagation within a vascular system. To evaluate the accuracy, most of the studies compared the simulation results with the outcome of physical experiments on a variety of phantom models, and only a limited number of studies have done face validity. Although a subset of the validated models is considered to be sufficiently accurate for the specific task for which they were developed and, therefore, are already being used in practice, these models are still under an ongoing development for improvement. Realism and computation time are two important requirements in catheter and guidewire modeling; however, the reviewed studies made a trade-off depending on the purpose of their model. Moreover, due to the complexity of the interaction with the vascular system, some assumptions have been made regarding the properties of both instruments and vascular system. Some validation studies have been reported but without a consistent experimental methodology.
BibTeX
@article{sharei2018navigation,
  author = {Hoda Sharei and Tanja Alderliesten and John J. van den Dobbelsteen and Jenny Dankelman},
  title = {{Navigation of guidewires and catheters in the body during intervention procedures: a review of computer-based models}},
  journal = {Journal of Medical Imaging},
  year = {2018},
  doi = {10.1117/1.jmi.5.1.010902},
}
Figure