TL;DR
Nonlinear optical microscopy techniques like SRS are powerful but can damage cells with their high-intensity pulsed lasers. This study systematically investigates the phototoxicity of these lasers on mouse Neuro2A cells, finding a surprising tolerance at typical imaging settings. The research establishes a threshold intensity (around 600 mW/µm²) above which significant phototoxicity, including cell damage and reactive oxygen species (ROS) generation, occurs, providing crucial guidance for experimental design.
What Problem Does This Paper Address?
Nonlinear optical microscopy (NLOM) techniques, such as two-photon microscopy and stimulated Raman scattering (SRS) microscopy, utilize pulsed lasers with high peak intensities. While these lasers enable advanced imaging capabilities, including label-free visualization of biological samples, their intense energy can perturb the native state of cells, leading to phototoxicity. This phototoxicity can manifest as cellular damage, altered gene expression, or changes in proliferation, potentially compromising downstream analyses, especially in experiments that pair imaging with molecular measurements like RNA sequencing. A critical gap exists in understanding the precise thresholds and mechanisms of this phototoxicity across various NLOM parameters, hindering the ability to design experiments that minimize cellular perturbation and ensure accurate interpretation of results. This paper aims to address this by systematically evaluating the phototoxic effects of SRS microscopy parameters on live cells.
Methods
The researchers employed a multimodal approach to assess phototoxicity:
- Cell Culture: Mouse Neuro2A (N2A) cells were cultured under standard conditions.
- Laser Exposure: Cells were exposed to various laser excitation settings relevant to SRS imaging. This involved tuning parameters such as average intensity, fluency (energy per unit area), pixel dwelling time, and the use of single or dual pulsed lasers. Specific exposure conditions were summarized in Table 1.
- Reactive Oxygen Species (ROS) Assay: The generation of ROS, a key indicator of cellular stress, was measured using the CellROX™ Green reagent. Increased fluorescence indicated ROS production.
- Proliferation Assay: Cell proliferation was monitored over time using automated phase-contrast imaging (IncuCyte ZOOM) to quantify changes in cell counts.
- Bulk RNA Sequencing (RNA-seq): Transcriptome-wide changes in gene expression were analyzed by extracting RNA from cells 1 hour after laser exposure and performing RNA-seq. Differential gene expression was assessed against negative control samples.
The study systematically varied laser parameters to identify thresholds for phototoxic effects and compared these to negative control groups.
Datasets
The study utilized cultured mouse Neuro2A (N2A) cells as the biological model. No new benchmark datasets were introduced.
Key Insights and Ideas
- Surprising Tolerance: N2A cells exhibit a remarkable tolerance to laser exposure at typical SRS imaging parameters. Many tested conditions did not lead to significant changes in proliferation or gene expression.
- Intensity is Key: The primary driver of phototoxicity appears to be laser intensity, rather than just fluency or pulse duration, at least within the tested ranges.
- Threshold for Damage: A clear threshold for significant phototoxicity was identified. Average intensities exceeding approximately 600 mW/µm² led to observable cell damage, decreased proliferation, and increased ROS generation.
- ROS as an Early Indicator: ROS generation serves as a sensitive, short-timescale indicator of cellular stress, appearing before observable changes in proliferation or gene expression.
- RNA-seq Robustness: Bulk RNA-seq, performed 1 hour post-exposure, did not reveal significant differential gene expression at many tested conditions, suggesting that transcriptional changes are either absent or occur on a longer timescale than captured.
- Guidance for Experimentation: The study provides concrete guidance for selecting laser parameters in SRS and other nonlinear optical microscopy techniques to minimize phototoxicity, crucial for paired imaging-molecular analysis experiments.
Key Results
- Laser exposure conditions with average intensities below ~600 mW/µm² (e.g., 47.72 mW/µm² and 238.58 mW/µm²) did not significantly affect N2A cell proliferation or gene expression over 30 hours post-exposure.
- Significant ROS generation was only observed at the highest tested average intensity (610.7 mW/µm²), which also resulted in a ~25% reduction in proliferation rate.
- At an average intensity of 610.7 mW/µm² with a fluency of 122 J/cm², cell burning was observed after repeated imaging frames.
- RNA-seq analysis showed no significant differential gene expression (FDR < 0.05) for most tested conditions, even those with higher fluency or near-burning intensity, suggesting robustness of the transcriptome to these perturbations.
- The study identified specific exposure conditions (e.g., I4-E5 at 610.7 mW/µm²) that induce phototoxicity, while others (e.g., I0-E0, I2-E2) are largely non-perturbative.
Limitations and Future Work
- Cell Type Specificity: The study focused on a single cell line (N2A). The observed tolerance might differ in other cell types, particularly primary cells or those with different metabolic profiles.
- Timescale of Gene Expression: While RNA-seq was performed 1 hour post-exposure, gene expression changes might occur on longer timescales not captured by this study.
- Repeated Exposure: The study primarily focused on single-pulse or single-scan exposure for most assays. While one condition (I1-E8) involved repeated imaging for ROS, a more comprehensive investigation into the cumulative effects of repeated laser exposure across all assays would be beneficial.
- ROS Scavengers and Temperature: The authors suggest future work exploring the use of ROS scavengers (e.g., N-Acetyl-L-cysteine) or lowering cell culture temperature to further mitigate photodamage, and to assess if these interventions impact gene expression or other cellular responses.
- Broader Parameter Space: The study explored a representative subset of SRS imaging parameters. Further investigation into other combinations of parameters, especially those used in specialized applications, could be valuable.
Summary updated