Abstract
This paper presents a comprehensive solution to overcome many of the fundamental challenges faced by current laparoscopic imaging systems. Our design features multiple tunable-focus microcameras integrated with a surgical port to provide panoramic intra-abdominal visualization with enhanced depth perception. Our system can be optically tuned to focus and zoom in on objects within a range of 3 mm to infinity, with a field of view adjustable between 36° and 130°. Our unique approach also eliminates the requirement of an exclusive imaging port and the need for navigation of cameras between ports during surgery.
Introduction
Standard minimally invasive surgery typically relies on a single, fixed-focus laparoscope inserted through a dedicated incision. This conventional approach suffers from several critical limitations: a restricted "soda straw" field of view (FOV), a lack of true depth perception, and the occupation of a surgical port solely for imaging. Furthermore, surgeons must frequently maneuver and clean the camera, while glare from reflective organ surfaces often obscures the surgical field.
To address these issues, we developed an integrated "smart" surgical port that deploys an array of four tunable-focus microcameras. This system provides a wide-angle panoramic view and enhanced visualization without requiring a dedicated imaging port.
System Design and Architecture
The Trocar-Camera Assembly (TCA)
The core of the system is the Trocar-Camera Assembly (TCA), which integrates optics directly onto the surgical trocar using a reconfigurable mechanical frame. This "umbrella-like" frame consists of four mechanical arms mounted at the distal end of the trocar. During insertion, the arms remain closed to fit through the initial incision. Once inside the abdominal cavity, they flare outward, positioning the microcameras around the periphery. This configuration ensures that the center of the port remains entirely unobstructed for the passage of surgical instruments.
Tunable Liquid Microlenses and Actuation
To achieve focusing without bulky mechanical motors, each microcamera utilizes a thermo-responsive hydrogel liquid lens. The lens consists of a water-oil meniscus within a microfluidic chamber. A ring of hydrogel acts as the actuator; when an integrated aluminum microheater warms the gel (between 25°C and 60°C), the hydrogel contracts. This contraction displaces water, changing the curvature of the meniscus and allowing the camera to focus across a wide range.
Optical Enhancements and Image Processing
To mitigate glare—a common problem in wet surgical environments—we fabricated and integrated thin-film Polyvinyl Alcohol (PVA) linear polarizers in front of the lenses. This hardware-based solution effectively blocks polarized light reflected from glossy tissue surfaces. To create the final visualization, images from the four cameras, which feature overlapping perspectives, are captured and stitched together to form a high-resolution panoramic view of up to 130°.
Results and Discussion
Experimental evaluation in a mimicked surgical environment demonstrated the system's robust capabilities:
- Focusing Range: The hydrogel-actuated lenses successfully achieved a tuning range from 3 mm to infinity.
- Field of View: The system provided an adjustable FOV between 36° and 130°, significantly exceeding the capabilities of standard laparoscopes.
- Glare Reduction: The integrated polarizers showed a dramatic qualitative reduction in "hot spots" on reflective surfaces, improving visual clarity.
- Panoramic Visualization: The system successfully produced a 128° panoramic image, providing a comprehensive view of the surgical field that aids in spatial awareness and depth perception.
Conclusion
The integrated TCA system represents a significant advancement in laparoscopic imaging. By combining multi-camera arrays, tunable liquid lenses, and hardware-based glare reduction, we have created a platform that enhances surgical visualization while freeing up valuable port access. Future work will focus on automating camera alignment, developing real-time 3D reconstruction algorithms, and transitioning toward wireless data transfer and medical-grade sterilization.
Summary updated