Abstract
We present a detailed theoretical analysis on image transmission via a single multimode fiber (MMF). A single MMF imaging model is developed to study the light wave propagation from the light source to the camera, by using free-space Fourier optics theory and mode-coupling theory. A mathematical expression is obtained for the complete single MMF imaging system, which is further validated by image-transmission simulations. Our model is believed to be the first theoretical model to describe the complete MMF imaging system based on the transmission of individual modes. Therefore, this model is robust and capable of analyzing MMF image transmission under specific mode-coupling conditions. We use our model to study bending-induced image blur in single-MMF image transmission, and the result has found a good agreement with that of existing experimental studies. These should provide important insights into future MMF imaging system developments.
Introduction: The "Hall of Mirrors" Challenge
Multimode fibers (MMFs) represent a potential breakthrough for ultra-thin endoscopy. Thinner than a human hair, a single MMF could theoretically deliver high-resolution images, replacing bulky traditional fiber bundles. However, MMFs act as a complex "hall of mirrors." Light traveling through the fiber spreads into thousands of spatial modes that interfere, resulting in a scrambled "speckle" pattern at the output.
While experimental "Transmission Matrices" (TMs) can unscramble these images, a robust theoretical model describing the entire process from first principles has been missing. This gap hinders our ability to predict how environmental factors, such as fiber bending, degrade image quality. This paper introduces a physical framework that explains why fiber bending scrambles images and reveals how fiber characteristics influence resilience to these distortions.
A Complete Theoretical Framework
To address this, researchers developed a comprehensive model combining two pillars of optical physics: free-space Fourier optics and mode-coupling theory. Unlike previous "black box" approaches, this model tracks the behavior of individual light modes—specifically Linear Polarized (LP) modes—throughout the entire system.
The framework accounts for the entire "closed-loop" pipeline: the illumination path (from source to object), the reflection off the target, and the return path through the fiber to the camera. By incorporating a mathematical description of bending-induced refractive index changes, the model can simulate how physical deformation alters mode coupling and reconfigures the Transmission Matrix.
Simulation and Validation
The model was validated using numerical simulations of a 1-meter-long step-index fiber with a 200 µm core. The primary simulation involved over 9,000 LP modes. Researchers simulated both the calibration phase (building the TM) and the imaging phase, using a "LIV" (Liverpool) letter pattern as the target object.
Crucially, the study explored the "Safety in Numbers" phenomenon. The model confirms that if the fiber's shape is known, the resulting image distortion is theoretically predictable. Furthermore, it was discovered that fibers supporting a higher number of modes are more resilient to bending-induced blur, as the distortion is "shared" across a larger mode population.
Key Results: Resilience and Reconstruction
The simulations successfully demonstrated that the model can "unscramble" speckle patterns to restore recognizable images. The study showed a direct correlation between bending radius and image blur: tighter bends lead to stronger mode coupling and more distorted reconstructions.
In a comparison between fibers with different mode counts, the version with more modes produced significantly clearer images at the same bending radius. This provides a critical theoretical basis for choosing high-mode-count fibers for flexible endoscopes to ensure better bending tolerance.
Limitations and Future Outlook
While robust, the current model assumes short fibers where polarization mixing is negligible and ignores radiative losses from extremely sharp bends. Furthermore, it addresses static bending rather than the dynamic, unknown deformations encountered in real-world medical applications. Future work aims to leverage this model to develop "smart" algorithms capable of real-time bending compensation, potentially eliminating the need for constant recalibration in clinical settings.
Summary updated