NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

Mid‐infrared dissolved CO2 measurement system using the ultra‐compact multipass cell

Hongda Chen, Weihao Dong, Yongyang Li, Mengru ZhangMicrowave 2022

Abstract

Abstract To measure natural gas hydrate, a mid‐infrared dissolved CO 2 measurement system is designed using the tunable diode laser spectroscopy with ultra‐compact multipass cell. The QCL laser of 4319 nm has a scanning range from 2315.05 to 2315.40 cm −1 , which contains 12 CO 2 and 13 CO 2 for isotope measurement and wide range of concentration detection. The optimized 3 f + 1 f wavelength modulation method reduces the limit of detection to 7.1 ppbv. An ultra‐compact multi‐pass cell with effective optical path length of 6.3 m and small volume of 15 ml is designed to achieve the system response time of 40 s. The dissolved CO 2 sensing system has been used in deep sea natural gas hydrate detection.
BibTeX
@inproceedings{chen2022midinfrared,
  author = {Hongda Chen and Chen Chen and Weihao Dong and Yongyang Li and Mengru Zhang},
  title = {{Mid‐infrared dissolved CO<sub>2</sub> measurement system using the ultra‐compact multipass cell}},
  booktitle = {Microwave},
  year = {2022},
  doi = {10.1002/mop.33450},
}

Loading related content...

Figure