Mid‐infrared dissolved CO2 measurement system using the ultra‐compact multipass cell
Abstract
Abstract To measure natural gas hydrate, a mid‐infrared dissolved CO 2 measurement system is designed using the tunable diode laser spectroscopy with ultra‐compact multipass cell. The QCL laser of 4319 nm has a scanning range from 2315.05 to 2315.40 cm −1 , which contains 12 CO 2 and 13 CO 2 for isotope measurement and wide range of concentration detection. The optimized 3 f + 1 f wavelength modulation method reduces the limit of detection to 7.1 ppbv. An ultra‐compact multi‐pass cell with effective optical path length of 6.3 m and small volume of 15 ml is designed to achieve the system response time of 40 s. The dissolved CO 2 sensing system has been used in deep sea natural gas hydrate detection.