Abstract
With the average age of population increasing in the world, the number of total hip replacement (THR) surgeries increases year by year. However, the failure rate of THR surgeries is up to 10% following primary surgery and 28% after revision arthroplasty, which has direct correlation to the implantation angles between prosthesis and pelvis. To improve the implanting accuracy of the prosthesis, we propose a measurement system and estimation algorithms. The system includes three parts: an initial attitude measurement instrument, which is used to determine the initial attitude of the anterior pelvic plane (APP), a real-time attitude measurement instrument, which aims to acquire the real-time attitude of the APP, and an acetabular cup angle measurement instrument, which obtains the relative angles between prosthesis and pelvis. Each of them is based on inertial measurement unit (IMU) and magnetometer. A quaternion-based extended Kalman filter is adopted to fuse the data from IMU and magnetometer to estimate the orientation, and the algorithm calculating the attitude of pelvis and relative angles between pelvis and prosthesis is designed in this paper. The reliability of this system, from the measurement point of view, has been verified by experiments. The experimental results show that the root-mean-square errors of APP attitude and acetabular cup implanting orientation are less than 1.6° and 3° with standard uncertainty of less than 0.22° and 0.17°, respectively. The accuracy of our system meets the requirement of the THR surgeries.
Introduction
Total Hip Replacement (THR) is a critical surgical intervention for patients with severe hip joint issues. Despite its success, failure rates remain significant—approximately 10% for primary surgeries and up to 28% for revisions. A primary cause of failure is dislocation, often resulting from the malpositioning of the acetabular cup. Surgeons typically aim for a "safe zone" (typically 5°–25° for anteversion and 30°–50° for inclination), but achieving this precision is challenging using traditional visual estimation or static preoperative plans, especially as the patient's pelvis moves during surgery. While CT-based navigation offers high accuracy, its high cost, increased surgical time, and radiation exposure limit its widespread adoption. This paper presents a real-time, radiation-free measurement system to track pelvic orientation and ensure accurate prosthesis implantation.
Methodology
The proposed system utilizes three specialized instruments integrated with inertial measurement units (IMUs) and magnetometers:
- Initial Attitude Measurement Instrument (IAMI): Used at the start of the procedure to define the Anterior Pelvic Plane (APP) by identifying three anatomical landmarks.
- Real-time Attitude Measurement Instrument (RAMI): A bone-mounted tracker that monitors pelvic movement throughout the surgery.
- Acetabular cup Angle Measurement Instrument (AAMI): Attached to the prosthesis inserter to measure the cup's orientation relative to the APP.
The system's core is a quaternion-based Extended Kalman Filter (EKF) that fuses data from 3-axis accelerometers, gyroscopes, and magnetometers. This approach leverages the high-frequency responsiveness of gyroscopes while correcting for drift using stable gravity and magnetic references. By establishing a fixed transformation matrix between the IAMI and RAMI, the system maintains a dynamic reference to the APP even when anatomical landmarks are obscured by surgical drapes.
Experimental Results
The system was evaluated using a physical pelvic phantom across seven different postures. Ground truth was established using high-precision mechanical instruments to verify the "true" attitude of the APP. The results demonstrate high accuracy and reliability:
- APP Tracking Accuracy: The Root-Mean-Square Error (RMSE) for the pelvic plane attitude was less than 1.6°.
- Implantation Accuracy: The RMSE for the final acetabular cup orientation was less than 3°.
- Precision: The standard uncertainty remained below 0.22°, indicating the system is highly repeatable and reliable for clinical use.
Discussion and Conclusion
The proposed IMU-based system provides a low-cost, high-accuracy alternative to traditional and CT-based navigation in THR. Its ability to track pelvic movement in real-time addresses a major limitation of static surgical planning. However, clinical success depends on the surgeon's precision in identifying anatomical landmarks during the initial setup. Future work will focus on clinical trials to evaluate performance in the presence of surgical tool interference and anatomical variability among patients.
Summary updated