Marker orientation in fiducial registration

Xiaofeng Liu, Hakan Çevıkalp, J. Michael FitzpatrickProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 2003 · DOI

Abstract

Fiducial markers are often employed in image-guided surgical procedures to provide positional information based on pre-operative images. In the standard technique, centroids of three or more markers are localized in both image space and physical space. The localized positions are used in a closed-form algorithm to determine the three-dimensional rigid-body transformation that will register the two spaces in the least-squares sense. In this work we present (1) a method for determining the orientation of the axis of symmetry of a cylindrical marker in a tomographic image and (2) an extension to the standard approach to rigid-body registration that utilizes the orientation of marker axes as an adjunct to the positions of their centroids. The extension is a closed-form, least-squares solution. Unlike the standard approach, the extension is capable of three-dimensional registration with only two markers. We evaluate the accuracy of the former method by means of CT and MR images of markers attached to a phantom and the accuracy of the latter method by means of computer simulations.
BibTeX
@inproceedings{liu2003marker,
  author = {Xiaofeng Liu and Hakan Çevıkalp and J. Michael Fitzpatrick},
  title = {{Marker orientation in fiducial registration}},
  booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
  year = {2003},
  doi = {10.1117/12.480860},
}