<title>Quasi-real-time x-ray microtomography system at the Advanced Photon Source</title>

Yuxin Wang, Francesco De Carlo, Ian Foster, Joseph A. Insley, Carl Kesselman, P. W. Lane, Gregor von Laszewski, Derrick C. Mancini, Ian McNulty, Mei-Hui Su, Brian TiemanProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE 1999 · DOI

TL;DR

This paper introduces a "quasi-realtime" X-ray microtomography system at the Advanced Photon Source that significantly speeds up the process of acquiring and reconstructing 3D images. By pipelining data acquisition, high-speed transfer, and parallel computation, they achieve near-instantaneous feedback on tomographic results, enabling interactive experimentation.

What Problem Does This Paper Address?

Traditional X-ray microtomography, while powerful for generating high-resolution 3D images, suffers from long acquisition and reconstruction times. This delay hinders interactive experimentation, making it difficult for researchers to quickly assess their sample, adjust parameters, or even confirm that the experiment is proceeding as expected. For user-shared facilities like synchrotron radiation sources, where beamtime is precious, this inefficiency is particularly problematic. The authors aim to bridge this gap by creating a system that provides rapid feedback, allowing for more efficient and dynamic experimental workflows.

Methods

The core of their solution is a highly integrated, pipelined system.

  • Data Acquisition: They use a straight-ray projection imaging scheme at the 2-BM beamline, employing a CdWO4 scintillator to convert X-rays to visible light, which is then imaged onto a CCD detector. This system can acquire projections at resolutions of 1-10 µm.
  • High-Speed Data Transfer: Immediately after each projection is acquired and saved locally, it's rapidly transferred via a 100 Mb/s network to a remote SGI Origin 2000 parallel computer.
  • Parallel Preprocessing and Reconstruction: The parallel computer, utilizing the Globus metacomputing toolkit, performs preprocessing (like white field removal, filtering, and alignment) and reconstruction concurrently with ongoing data acquisition.
  • Visualization: The reconstructed data is then sent to a visualization computer (Viz Engine) for volume rendering. This allows operators at the beamline to view rendered images within minutes of data acquisition completion.
  • Data Format: They use the Hierarchical Data Format (HDF) with NeXus conventions for organizing experimental data, ensuring platform independence and self-description.

Datasets

The paper describes experiments using two types of samples:

  • A micromachined gear (approximately 100 µm diameter), acquired using 15 keV X-rays in phase contrast.
  • A carpenter ant's head, acquired using 9 keV X-rays in absorption-contrast mode.

These are not presented as benchmark datasets but rather as examples demonstrating the system's capabilities.

Key Insights and Ideas

  • Pipelining is Key: The most significant insight is that by tightly integrating acquisition, transfer, and parallel processing, the overall time to get a reconstructed 3D image can be drastically reduced, approaching the acquisition time itself.
  • "Quasi-Realtime" is Sufficient: For many experimental scenarios, achieving results within minutes, rather than seconds, is sufficient to enable interactive adjustments and feedback, revolutionizing experimental efficiency.
  • Leveraging Supercomputing: The paper highlights the critical role of parallel computing resources in handling the massive data volumes generated by modern X-ray sources for rapid reconstruction.
  • Data Format Matters for Integration: The choice of HDF and NeXus conventions facilitates the seamless transfer and interpretation of data across different computational stages.

Key Results

  • A complete tomographic dataset can be collected in approximately 15 minutes.
  • Reconstruction results are available for viewing minutes after data acquisition is complete.
  • For test runs with 180 projections (512x512 pixels), reconstruction on 20 out of 128 processors took less than 10 minutes.
  • When acquiring data in cycles (e.g., 12-24 projections), reconstruction results can be viewed at the experiment station within 2-3 minutes.
  • The system currently achieves resolutions of 1-10 µm, with plans to reach below 100 nm.

Limitations and Future Work

  • Resolution Limits: Current resolution is around 1-10 µm, with plans to improve this to sub-100 nm by using focused beams and advanced optics.
  • Acquisition Time: While significantly reduced, acquisition time is still a bottleneck. Future work aims to reduce it to seconds by improving detectors and using more efficient monochromators.
  • Computation Time: Computation time is also being targeted for reduction to under 1 minute through software optimization and parallelization.
  • Subsampling for Visualization: The Viz Engine has a data volume size limitation (256x256x256 pixels), requiring subsampling of larger reconstructions for real-time visualization.
  • User Interface: A user-friendly graphical user interface (GUI) is under development to allow general users to control the entire pipeline interactively from the beamline.
  • Detector Improvements: A scintillator better matched for 5-20 keV X-ray absorption and a high numerical aperture objective are needed to improve efficiency and resolution.

Summary updated

Abstract

The combination of high-brilliance x-ray sources, fast detector systems, wide-bandwidth networks, and parallel computers can substantially reduce the time required to acquire, reconstruct, and visualize high-resolution three- dimensional tomographic data sets. A quasi-realtime computed x-ray microtomography system has been implemented at the 2-BM beamline at the Advanced Photon Source at Argonne National Laboratory. With this system, a complete tomographic data set can be collected in about 15 minutes. Immediately after each projection is obtained, it is rapidly transferred to the Mathematics and Computing Sciences Division where preprocessing and reconstruction calculations are performed concurrently with the data acquisition by a SGI parallel computer. The reconstruction results, once completed, are transferred to a visualization computer that performs the volume rendering calculations. Rendered images of the reconstructed data are available for viewing back at the beamline experiment station minutes after the data acquisition was complete. The fully pipelined data acquisition and reconstruction system also gives us the option to acquire the tomographic data set in several cycles, initially with coarse then with fine angular steps. At present the projections are acquired with a straight-ray projection imaging scheme using 5 - 20 keV hard x rays in either phase or amplitude contrast mode at a 1 - 10 micrometer resolution. In the future, we expect to increase the resolution of the projections to below 100 nm by using a focused x-ray beam at the 2-ID-B beamline and to reduce the combined acquisition and computation time to the 1 min scale with improvements in the detectors, network links, software pipeline, and computation algorithms.
BibTeX
@inproceedings{wang1999lt,
  author = {Yuxin Wang and Francesco De Carlo and Ian Foster and Joseph A. Insley and Carl Kesselman and P. W. Lane and Gregor von Laszewski and Derrick C. Mancini and Ian McNulty and Mei-Hui Su and Brian Tieman},
  title = {{\<title\>Quasi-real-time x-ray microtomography system at the Advanced Photon Source\</title\>}},
  booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
  year = {1999},
  doi = {10.1117/12.363735},
}
Figure