Long-term performance of intracortical microelectrode arrays in 14 BrainGate clinical trial participants

Nick Hahn, Elias A. Stein, John P. Donoghue, John D. Simeral, Leigh R. Hochberg, Francis R. WillettmedRxiv 2025

Abstract

Brain-computer interfaces have enabled people with paralysis to control computer cursors, operate prosthetic limbs, and communicate through handwriting, speech, and typing. Most high-performance demonstrations have used silicon microelectrode "Utah" arrays to record brain activity at single neuron resolution. However, reports so far have typically been limited to one or two individuals, with no systematic assessment of the longevity, decoding accuracy, and day-to-day stability properties of chronically implanted Utah arrays. Here, we present a comprehensive evaluation of 20 years of neural data from the BrainGate and BrainGate2 pilot clinical trials. This dataset spans 2,319 recording sessions and 20 arrays from the first 14 participants in these trials. On average, arrays successfully recorded neural spiking waveforms on 35.6% of electrodes, with only a 7% decline over the study enrollment period (up to 7.6 years, with a mean of 2.8 years). We assessed movement intention decoding performance using a "decoding signal-to-noise ratio" (dSNR) metric, and found that 11 of 14 arrays provided meaningful movement decoding throughout study enrollment (dSNR > 1). Three arrays reached a peak dSNR greater than 4.5, approaching that achieved during able-bodied computer mouse control (6.29). We also found that dSNR increases logarithmically with the number of electrodes, providing a pathway for scaling performance. Longevity and reliability of Utah array recordings in this study were better than in prior nonhuman primate studies. However, achieving peak performance consistently will require addressing unknown sources of variability.

Introduction

While Brain-Computer Interfaces (BCIs) have made headlines for allowing people with paralysis to type or control robotic limbs, most high-performance results come from "heroic" case studies involving only one or two participants. This has left a major gap in our understanding of how these devices perform across a diverse population over many years.

Engineers and clinicians need to know: Do the electrodes inevitably fail due to the brain's immune response? How much does decoding accuracy vary between people? And how stable is the neural "language" of movement intention from one day to the next? By analyzing two decades of data, this study moves BCI research from isolated demonstrations toward a systematic engineering assessment of device longevity and reliability.

Methods

The researchers curated two primary datasets from 2,319 recording sessions: a "yield dataset" to measure how many electrodes remained active, and a "decoding dataset" to measure functional performance.

The Utah microelectrode array shown next to a dime for scale.

Key technical approaches included:

  • Decoding Signal-to-Noise Ratio (dSNR): Instead of using task-specific metrics (like words per minute), the authors used dSNR. This vector-based metric quantifies how much movement information is present in the neural population relative to noise, making it independent of the specific decoding algorithm or task used at the time.
  • Spike Band Power (SBP): Rather than just looking for discrete "spikes" from single neurons, they utilized power in the 250–5000 Hz band. This captures low-amplitude signals that are often missed by traditional thresholding but are highly informative for decoding.
  • Subsampling & Extrapolation: To understand the future of the field, they randomly subsampled electrodes to model how performance scales with channel count, eventually extrapolating to 1024-channel devices currently in development by companies like Neuralink and Paradromics.

Datasets and Study Scope

The study utilizes a massive, longitudinal dataset from the BrainGate and BrainGate2 pilot clinical trials (2004–2019).

  • Participants: 14 individuals with paralysis (e.g., ALS, spinal cord injury).
  • Hardware: 20 chronically implanted 96-channel "Utah" arrays.
  • Scope: 2,319 total recording sessions and 634 specific decoding sessions.
  • Benchmark: The study also recorded cursor movements from 9 able-bodied volunteers using a standard computer mouse to establish a "gold standard" dSNR of 6.29.
Study overview showing participant setup (d), cursor trajectories (e), and longitudinal data collection across 14 participants (f).

Key Insights and Results

  • Longevity is better than feared: Contrary to some non-human primate studies suggesting rapid decline, human arrays showed only a 7% average decline in yield over years. Some arrays remained functional for over seven years.
  • Yield vs. Performance: An array can have a low "yield" (few clear individual neurons) but still provide high-quality decoding. This is because "background" neural activity (Spike Band Power) contains significant information about intent, even if the signal is too small to isolate a single cell.
  • The Logarithmic Scaling Law: dSNR increases logarithmically with the number of electrodes. This provides a mathematical "North Star" for the industry: to reach the performance of an able-bodied mouse user, we need significantly more channels.
  • Functional Persistence: 11 of 14 arrays provided meaningful movement decoding (dSNR > 1) throughout the entire study period. Three arrays reached a peak dSNR > 4.5, which is roughly 70% of the performance achieved by able-bodied mouse users.
  • Short-term Stability: Neural tuning remains stable over roughly 30 days (correlation > 0.6), suggesting BCIs could function for weeks without daily re-calibration.

Discussion and Future Work

The authors acknowledge significant unexplained variability. While some arrays performed beautifully for years, others failed early or had consistently low signals. The causes—ranging from surgical technique and implant location to individual disease states and the body’s foreign body response—remain unknown sources of variability that must be solved to make BCIs a reliable consumer technology.

Future work will likely focus on "hot spot" targeting (finding the exact best millimeters of the motor cortex to implant) and the transition to higher-channel-count, fully implanted wireless systems that can leverage the logarithmic scaling identified here.

Summary updated

Abstract

Brain-computer interfaces have enabled people with paralysis to control computer cursors, operate prosthetic limbs, and communicate through handwriting, speech, and typing. Most high-performance demonstrations have used silicon microelectrode "Utah" arrays to record brain activity at single neuron resolution. However, reports so far have typically been limited to one or two individuals, with no systematic assessment of the longevity, decoding accuracy, and day-to-day stability properties of chronically implanted Utah arrays. Here, we present a comprehensive evaluation of 20 years of neural data from the BrainGate and BrainGate2 pilot clinical trials. This dataset spans 2,319 recording sessions and 20 arrays from the first 14 participants in these trials. On average, arrays successfully recorded neural spiking waveforms on 35.6% of electrodes, with only a 7% decline over the study enrollment period (up to 7.6 years, with a mean of 2.8 years). We assessed movement intention decoding performance using a "decoding signal-to-noise ratio" (dSNR) metric, and found that 11 of 14 arrays provided meaningful movement decoding throughout study enrollment (dSNR > 1). Three arrays reached a peak dSNR greater than 4.5, approaching that achieved during able-bodied computer mouse control (6.29). We also found that dSNR increases logarithmically with the number of electrodes, providing a pathway for scaling performance. Longevity and reliability of Utah array recordings in this study were better than in prior nonhuman primate studies. However, achieving peak performance consistently will require addressing unknown sources of variability.
BibTeX
@misc{hahn2025long,
  author = {Nick Hahn and Elias A. Stein and John P. Donoghue and John D. Simeral and Leigh R. Hochberg and Francis R. Willett},
  title = {{Long-term performance of intracortical microelectrode arrays in 14 BrainGate clinical trial participants}},
  year = {2025},
  doi = {10.1101/2025.07.02.25330310},
  howpublished = {medRxiv},
}
Figure