Local threshold for segmented attenuation correction of PET imaging of the thorax
Lists
Only you see this
Abstract
A local threshold for segmented attenuation correction technique has been developed for positron emission tomography using short (2-3 minutes) post-injection transmission scans. The technique implements an optimal threshold method on localized histograms to get pseudo-anatomic segmentation on transmission images. Theoretical values of attenuation coefficients are assigned to corresponding anatomic regions. Emission images are reconstructed using attenuation correction factors computed by forward-projecting segmented transmission images. Phantoms and clinical cardiac images are studied using this technique. The technique corrects emission images with accuracy similar to the standard, pre-injection method, reduces noise in the corrected emission images, and offers the potential for increased patient throughput by enabling shorter, post-injection data acquisition.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
BibTeX
@article{xu1994local,
author = {Mai Xu and W.K. Luk and Paul Cutler and W.M. Digby},
title = {{Local threshold for segmented attenuation correction of PET imaging of the thorax}},
journal = {IEEE Transactions on Nuclear Science},
year = {1994},
doi = {10.1109/23.322944},
}Loading related content...