Abstract
The Large High-Altitude Air Shower Observatory (LHAASO) KM2A array utilizes 1,171 Muon Detectors (MDs), each consisting of a water Cherenkov detector buried under 2.5 meters of soil. To ensure performance before permanent burial, an on-site test system was developed. This system, featuring a scintillator telescope and a portable data acquisition unit, evaluates water quality and liner reflectivity by measuring charge and time resolutions.
The Challenge of Underground Muon Detection
LHAASO’s KM2A array relies on Muon Detectors to distinguish high-energy gamma rays from the background noise of hadronic cosmic rays. These detectors are massive 44-ton water tanks. To shield them from low-energy electromagnetic particles, they are buried under a 2.5-meter thick layer of soil.
Once buried, maintenance is virtually impossible. Contaminated water or degraded reflective liners would permanently reduce detector sensitivity. Therefore, a reliable on-site testing protocol is required to verify each unit's integrity during the construction phase.
On-Site Test System Methodology
The research team developed two primary iterations of the test system to validate MD performance:
1. The Scintillator Telescope
The initial test setup employed a telescope composed of two vertically stacked scintillator detectors, each shielded by 5 cm of lead. By requiring a coincidence signal between the two scintillators, the system filters out low-energy background radiation. Simulations indicate that this configuration achieves a muon selection purity of approximately 87%, providing a reliable "standard candle" for calibrating the water tank's response.
2. The Portable "Suitcase" System
To accommodate the rapid pace of construction, the electronics were miniaturized into a portable, remote-controlled "suitcase" unit. This streamlined Data Acquisition (DAQ) system allows for efficient testing without the bulky telescope hardware, enabling remote operation and faster deployment across the vast array site.
Key Performance Metrics
The system evaluates two critical parameters:
- Number of Photoelectrons (NPE): Quantifies the total light collected by the photomultiplier tube (PMT).
- Decay Time Constant of Signals (DTCS): Measures how long Cherenkov light persists within the tank.
Analysis revealed a strong correlation between DTCS and NPE. Longer decay times indicate higher water purity and better liner reflectivity, resulting in a stronger overall signal. This insight led to the optimization of the internal lining, where doubling the Tyvek layers significantly improved performance.
Results and Deployment
As of the project's early stages, the system successfully validated the first 247 MDs. The lead-shielded telescope confirmed high muon purity, and the data showed excellent consistency across manufactured units. These tests ensure that the detectors will remain functional throughout their projected 20-year lifespan, providing a baseline for long-term health monitoring.
Summary updated