LHAASO KM2A Muon Detector On-site Test System

S. P. Zhao, Yi Zhang, Gang Xiao, Hongliang Dai, Guang-Guang Xin, Xiong Zuo, Zhe Li, Cong Li, Lingyu Wang, Hui Wang, Ning ChengProceedings of 36th International Cosmic Ray Conference — PoS(ICRC2019) 2019 · DOI

Abstract

The Large High-Altitude Air Shower Observatory (LHAASO) KM2A array utilizes 1,171 Muon Detectors (MDs), each consisting of a water Cherenkov detector buried under 2.5 meters of soil. To ensure performance before permanent burial, an on-site test system was developed. This system, featuring a scintillator telescope and a portable data acquisition unit, evaluates water quality and liner reflectivity by measuring charge and time resolutions.

The Challenge of Underground Muon Detection

LHAASO’s KM2A array relies on Muon Detectors to distinguish high-energy gamma rays from the background noise of hadronic cosmic rays. These detectors are massive 44-ton water tanks. To shield them from low-energy electromagnetic particles, they are buried under a 2.5-meter thick layer of soil.

Schematic of the LHAASO Muon Detector (MD) buried under 2.5m of soil overburden.

Once buried, maintenance is virtually impossible. Contaminated water or degraded reflective liners would permanently reduce detector sensitivity. Therefore, a reliable on-site testing protocol is required to verify each unit's integrity during the construction phase.

On-Site Test System Methodology

The research team developed two primary iterations of the test system to validate MD performance:

1. The Scintillator Telescope

The initial test setup employed a telescope composed of two vertically stacked scintillator detectors, each shielded by 5 cm of lead. By requiring a coincidence signal between the two scintillators, the system filters out low-energy background radiation. Simulations indicate that this configuration achieves a muon selection purity of approximately 87%, providing a reliable "standard candle" for calibrating the water tank's response.

Schematic and physical components of the scintillator telescope test system.

2. The Portable "Suitcase" System

To accommodate the rapid pace of construction, the electronics were miniaturized into a portable, remote-controlled "suitcase" unit. This streamlined Data Acquisition (DAQ) system allows for efficient testing without the bulky telescope hardware, enabling remote operation and faster deployment across the vast array site.

Key Performance Metrics

The system evaluates two critical parameters:

  • Number of Photoelectrons (NPE): Quantifies the total light collected by the photomultiplier tube (PMT).
  • Decay Time Constant of Signals (DTCS): Measures how long Cherenkov light persists within the tank.

Analysis revealed a strong correlation between DTCS and NPE. Longer decay times indicate higher water purity and better liner reflectivity, resulting in a stronger overall signal. This insight led to the optimization of the internal lining, where doubling the Tyvek layers significantly improved performance.

Results and Deployment

As of the project's early stages, the system successfully validated the first 247 MDs. The lead-shielded telescope confirmed high muon purity, and the data showed excellent consistency across manufactured units. These tests ensure that the detectors will remain functional throughout their projected 20-year lifespan, providing a baseline for long-term health monitoring.

Summary updated

Abstract

The Muon detector (MD) of LHAASO-KM2A (Large High-Altitude Air Shower Observatory 1km2 Array) [1] is a water Cherenkov detector with an effective area of ~36 m2. The 2.5 m thick overburden soil is used as a shield to prevent low energy electromagnetic particles from reaching water. Each MD must be well tested before the covering process of the thick soil. In this case, the on-site test system is designed and manufactured, consisting of a telescope and a data acquisition system. The telescope is a combination of two vertically placed scintillator detectors covered by 4cm thick lead each. Charge resolution and time resolution are estimated by analysing the test data. In addition, waveforms of the MD signal are recorded each second, which will serve to evaluate the water quality of MD.
BibTeX
@inproceedings{zhao2019lhaaso,
  author = {S. P. Zhao and Yi Zhang and Gang Xiao and Hongliang Dai and Guang-Guang Xin and Xiong Zuo and Zhe Li and Cong Li and Lingyu Wang and Hui Wang and Ning Cheng},
  title = {{LHAASO KM2A Muon Detector On-site Test System}},