Left Atrial scar assessment using imaging with isotropic spatial resolution and compressed sensing

Mehmet Akçakaya, Susie N. Hong, Raymond H. Chan, Tamer Basha, Mehdi H. Moghari, Kraig V. Kissinger, Beth Goddu, Mark E. Josephson, Warren J. Manning, Reza NezafatJournal of Cardiovascular Magnetic Resonance 2012 · DOI

Abstract

We assess left atrial scar using late gadolinium enhancement (LGE) with isotropic spatial resolution of 1.4mm³ by using highly accelerated Likelihood-based Iterative Thresholding (LOST) reconstruction.

Introduction

Atrial fibrillation (AF) is a prevalent heart rhythm disorder frequently managed through Pulmonary Vein Isolation (PVI). This procedure utilizes radiofrequency energy to create scars that obstruct irregular electrical signals. Evaluating the success of PVI relies on Late Gadolinium Enhancement (LGE) MRI to visualize these ablation scars.

However, conventional LGE MRI typically employs "non-isotropic" imaging. While providing high resolution in the axial plane, these scans use thick slices (e.g., 4mm). Given the complex 3D geometry of the left atrium and pulmonary veins, these thick slices result in significant blurring when viewed from sagittal or coronal orientations, complicating precise scar mapping.

Methodology

This study compared two imaging protocols across a clinical cohort of 23 patients (average age 58.1, 6 females). The cohort included 9 pre-procedure patients, 8 post-procedure patients, and 3 scanned both before and after. The protocols were:

  1. Standard LGE: Non-isotropic acquisition (1.4 x 1.4 x 4.0 mm³).
  2. High-Resolution Isotropic LGE: Acquisition with equal resolution in all directions (1.4 x 1.4 x 1.4 mm³).

To achieve high-resolution isotropic imaging without a threefold increase in scan time, researchers employed Compressed Sensing. Specifically, they utilized:

  • Undersampling: Only approximately 1/3 of the raw k-space data was collected, maintaining full sampling at the center while randomly discarding peripheral data.
  • LOST Reconstruction: The Likelihood-based Iterative Thresholding (LOST) technique was used to reconstruct the missing data.
  • Acquisition Parameters: Both protocols required approximately 4 minutes, employing ECG-triggering and navigator-gating to mitigate motion artifacts from heartbeats and respiration.

Results and Discussion

The study successfully produced diagnostic images in 20 out of 23 cases. LGE was detected in 14 patients via standard scans and 12 patients via LOST-reconstructed scans, with disagreement in only two cases.

The primary advantage of isotropic imaging lies in its consistency across orientations. Because voxels are cubic (1.4mm³), images can be reformatted into any plane without the "stair-stepping" artifacts characteristic of standard non-isotropic scans.

Comparison of non-isotropic (top) and isotropic (bottom) LGE MRI scans in axial (a), sagittal (b), and coronal (c) views. Arrows highlight regions of enhancement.

Quantitative image quality scoring (on a 4-point scale) revealed a trade-off:

  • Axial View: Standard scans (3.76) outperformed LOST (2.90).
  • Sagittal View: LOST (3.33) significantly outperformed Standard (2.48).
  • Coronal View: LOST (3.00) significantly outperformed Standard (2.24).

While standard scans maintain higher sharpness in their primary acquisition plane (axial), the isotropic LOST scans provide superior multi-angle clarity, which is critical for mapping 3D structures like the pulmonary veins. Qualitative findings suggest that LOST-reconstructed images allow for the visualization of separate, distinct regions of enhancement that appear as a single blurred blob in standard scans.

Conclusion and Future Work

This study demonstrates that 3-fold acceleration via compressed sensing allows for a trade-off between imaging speed and spatial resolution, enabling isotropic LGE MRI. While the LOST reconstruction may introduce minor smoothing in the axial plane, it offers significantly better visualization in sagittal and coronal views.

Future research with larger cohorts is necessary to validate these findings and determine if the improved multi-planar resolution translates into enhanced clinical outcomes or more accurate diagnostic capabilities for physicians.

Summary updated

Abstract

We assess left atrial scar using late gadolinium enhancement (LGE) with isotropic spatial resolution of 1.4mm 3 by using highly accelerated LOST [1] reconstruction.
BibTeX
@article{akcakaya2012left,
  author = {Mehmet Akçakaya and Susie N. Hong and Raymond H. Chan and Tamer Basha and Mehdi H. Moghari and Kraig V. Kissinger and Beth Goddu and Mark E. Josephson and Warren J. Manning and Reza Nezafat},
  title = {{Left Atrial scar assessment using imaging with isotropic spatial resolution and compressed sensing}},
  journal = {Journal of Cardiovascular Magnetic Resonance},
  year = {2012},
  doi = {10.1186/1532-429x-14-s1-o8},
}
Figure