TL;DR
This paper introduces and evaluates an automatic approach for classifying laryngeal lesions based on vascular patterns observed in Contact Endoscopy and Narrow Band Imaging (CE-NBI). The automatic method shows promise in reducing subjective interpretation errors common in manual classification by clinicians, potentially serving as a valuable assistant tool.
What Problem Does This Paper Address?
Laryngeal cancer is a significant health concern, and early diagnosis is crucial for better patient outcomes and preserving vocal function. Specific changes in the blood vessel patterns within the laryngeal mucosa, particularly on the vocal folds, are indicative of benign or malignant lesions. Contact Endoscopy combined with Narrow Band Imaging (CE-NBI) offers a way to visualize these vascular changes in real-time during surgery. However, visually interpreting these patterns is challenging for clinicians due to similarities between benign and malignant lesions, leading to subjective evaluations and potential misclassifications. This paper aims to bridge this gap by developing and comparing an automatic classification system against manual expert evaluation.
Methods
The study compares two approaches: a manual classification by otolaryngologists and an automatic classification using a machine learning algorithm.
Manual Approach: Six otolaryngologists (three specialists and three residents) independently reviewed CE-NBI images from a subset of patients. They were blinded to the histological diagnosis and used the European Laryngological Society (ELS) guideline to classify lesions based on Perpendicular Vascular Changes (PVC).
Automatic Approach: An updated version of a previously proposed algorithm was used. This involved:
- Pre-processing: Vessel enhancement using the Jerman filter (an improvement over the Frangi filter for natural variations) and segmentation.
- Feature Extraction: Extracting 24 geometrical features related to gradient direction and curvature.
- Supervised Classification: Employing four classifiers: Support Vector Machine (SVM) with Polykernel and Radial Basis Function (RBF) kernels, k-Nearest Neighbor (kNN), and Random Forest Classifier (RFC). Hyperparameter tuning for these classifiers was performed using grid search and 10-fold cross-validation to optimize performance on the specific dataset.
Datasets
The study utilized a CE-NBI dataset comprising 1632 images from 68 patients. Histological diagnoses, based on the WHO classification, were used to label the images.
- Subset I: Approximately 20% of the dataset (336 images), randomly selected from each patient, was used for manual evaluation by clinicians and as the testing set for the automatic approach.
- Subset II: The remaining 80% of the dataset (1296 images) was used as the training set for the automatic approach.
Key Insights and Ideas
- Subjectivity of Manual Interpretation: The study strongly reinforces that manual interpretation of vascular patterns in CE-NBI images is highly subjective and significantly influenced by the clinician's experience. Specialists performed better than residents.
- Automatic Approach as a Decision Support: The automatic classification system demonstrates the potential to act as an assistant, particularly in cases of disagreement among clinicians or when all clinicians misclassify a lesion.
- Overcoming Visual Similarities: The automatic approach, by quantifying geometric features, can better differentiate subtle vascular pattern differences that are visually challenging for humans, especially between certain benign (e.g., Papillomatosis) and malignant lesions.
- Histopathology Matters: The performance of both manual and automatic methods varies significantly depending on the specific histopathology of the lesion, highlighting the complexity of certain diagnoses.
Key Results
- Manual vs. Automatic Performance: While specialists achieved a sensitivity of 0.955, their specificity was lower (0.727) compared to the automatic classifiers. The automatic approach, particularly SVM with RBF, achieved high specificity (0.981) and good sensitivity (0.806). RFC showed the highest sensitivity (0.846) with good specificity (0.895).
- Reduced Misclassification: The automatic approach significantly reduced misclassification rates for challenging histopathologies like Papillomatosis and Mild Dysplasia, where manual classification struggled.
- SVM with RBF Excellence: The SVM with RBF kernel demonstrated the best overall performance, showing no misclassification per patient across all histopathology categories in the study.
- Agreement/Disagreement Analysis: Figure 2 visually depicts the agreement and disagreement between clinicians and classifiers. While individual images could be misclassified by classifiers, the patient-level classification by the automatic approach was often correct, especially in Category I where all clinicians agreed.
Limitations and Future Work
- Dataset Imbalance: The CE-NBI dataset had more images for benign lesions than malignant ones, potentially affecting the classifiers' performance on malignant cases.
- Feature Set Limitations: The 24 extracted features primarily focus on geometrical characteristics of vascular patterns. Future work should explore incorporating other relevant characteristics for improved classification.
- Differentiation of PVC Angles: The paper suggests developing new methods to better differentiate between wide and narrow-angled points of PVCs, which is a source of diagnostic difficulty.
- Balancing Datasets: Future research should aim to balance the number of images for benign and malignant lesions in the dataset.
- Expanding Feature Set: Incorporating additional features beyond geometry is recommended.
- Assistant System Development: The authors envision further development of the automatic approach as an assistant system to aid clinicians in treatment planning and intraoperative decision-making.